MoblVac Portable Wound Care Pump; an AC/DC powered portable aspirator that provides negative pressure for the...
Ohio Medical Corporation began a recall of MoblVac Portable Wound Care Pump; an AC/DC powered portable aspirator that provides negative pressure for the application of wound drainage; Ohio Medical... on . The reason FDA records is "Ohio Medical has received a report of an explosive event with a MoblVac Portable Woundcare Pump housing where parts were separated while in an enclosed nylon transport bag during use". As of , the recall status on record is Terminated.
Recall record
- Recall number
- Z-1777-2010
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Ohio Medical LLC
- Product
- MoblVac Portable Wound Care Pump; an AC/DC powered portable aspirator that provides negative pressure for the application of wound drainage; Ohio Medical Corporation, Gurnee, IL 60031; Models 769600 (with Truseal), 759600 (refurbished), 769600 Rental, 769601 (rental), 769602 (evaluation pump with Truseal) and 769610 (sales unit with case).
- Product code
- BTA Pump, Portable, Aspiration (Manual Or Powered)
- Reason for recall
- Ohio Medical has received a report of an explosive event with a MoblVac Portable Woundcare Pump housing where parts were separated while in an enclosed nylon transport bag during use.
- FDA root cause
- Other
- Action
- Ohio Medical sent Urgent Medical Device Recall letters dated 5/7/10 to the direct accounts on 5/10/10, with follow-up telephone calls on 5/11/10. The accounts were informed of the one report of an explosion of the pump, and were requested to take the pumps out of service and return the pumps to Ohio Medical. The accounts were requested to complete and fax back to Ohio Medical the enclosed acknowledgement form indicating the number of pumps being returned. Any questions were directed to Anita Christ at 847-855-6270. Follow-up letters dated 5/13/10 were sent to the accounts on 5/14/10, providing instructions for the return of the pumps and bags. Shipping boxes will be sent to the accounts based on the number of pumps reported on the acknowledgment form.
- Quantity in commerce
- 375 units
- Distribution
- Nationwide and internationally to Qatar and Oman
- Product codes and lots
- Serial numbers NPJ0001 through NPJ0348 and NPK0001 through NPK0095.