Syngo Imaging Version Lines VB20, VB30, VB35 in combination with a RIS
Siemens Medical Solutions USA, Inc began a recall of Syngo Imaging Version Lines VB20, VB30, VB35 in combination with a RIS on . The reason FDA records is "Siemens has become aware of a potential malfunction and hazard to patients when using Syngo Imaging version lines VB20, VB30, VB35A in combination with a RIS that is violating DICOM Standard by creating non-unique study...". As of , the recall status on record is Terminated, and MDM tags the root cause as design.
Recall record
- Recall number
- Z-1892-2010
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Siemens Medical Solutions USA, Inc.
- Product
- Syngo Imaging Version Lines VB20, VB30, VB35 in combination with a RIS.
- Product code
- LLZ System, Image Processing, Radiological
- Reason for recall
- Siemens has become aware of a potential malfunction and hazard to patients when using Syngo Imaging version lines VB20, VB30, VB35A in combination with a RIS that is violating DICOM Standard by creating non-unique study instance UIDs.
- Root cause tag
- design
- FDA root cause
- Device Design
- Action
- The recalling firm sent a Customer Safety Advisory Notice, dated 04-06-2010, to customers explaining when the malfunction occurs and the potential risks, steps to avoid potential risk of the issue by ensuring that the RIS works according to DICOM Standard and does not create double study instance UIDs. Siemens Service Reps will contact the consignees in a timely manner to check if there are symptoms for the reported issues. The issue will be resolved with a Software-Update for each affected version, except for version VB20. All customers on VB20 will be upgraded.
- Quantity in commerce
- 3 units
- Distribution
- Nationwide Distribution -- Including CT, NM, & WI.
- Product codes and lots
- Material number 10014063. Serial numbers 1043, 1114, and 1198. Version Lines VB20, VB30, & VB35.