Romedic, Inc. began a recall of EVA 450 Patient Lift Battery Charger on . The reason FDA records is "Incorrect battery chargers were supplied with the EVA 450 Patient Lifts. Charger unit stops working and will not charge lift. New chargers were supplied to customers". As of , the recall status on record is Terminated.
Recall record
- Recall number
- Z-1736-2010
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Romedic, Inc.
- Product
- EVA 450 Patient Lift Battery Charger.
- Product code
- FSA Lift, Patient, Non-Ac-Powered
- Reason for recall
- Incorrect battery chargers were supplied with the EVA 450 Patient Lifts. Charger unit stops working and will not charge lift. New chargers were supplied to customers.
- FDA root cause
- Other
- Action
- The recalling firm sent out a letter with instructions as follows: 1) Locate EVA 450 Patient Lift(s), 2) Fill out Customer Response Form and fax to 888-691-1091, 3) Replace battery charger(s) with new battery charger(s) supplied with this notification; NEW battery charger to be used according to EVA 450 Instructions, 4) Repackage replaced battery charger(s) in original packing in which shipment was recieved, 5) Apply supplied sticker "EVA 450 Battery Charger RECALL" on outside of box, 6) Apply supplied UPS return label on outside of box and contact UPS for pickup and mail to the following Address: RoMedic, Inc., 300 West Chestnut St., Ephrata, PA 17522; Attn: EVA 450 Battery Charger Recall. Additional questions are directed to the company at phone: 888-625-4343.
- Quantity in commerce
- 76 units
- Distribution
- Nationwide distribution.
- Product codes and lots
- Serial numbers: 373317696102209, 373317696102210, 373317696102224, 373317696102227, 373317696102228, 373317696102232, 373317696102235, 373317696102242, and 373317696102244.