Z-2236-2010
Olympus SD Snares, Reusable Electrosurgical Snares. Models: SD-9L/U-1, SD-11L/U-1, SD-12L/U-1, SD-13L/U-1...
OLYMPUS Corporation of the Americas / initiated 2010-05-03 / Terminated
Olympus America Inc. began a recall of Olympus SD Snares, Reusable Electrosurgical Snares. Models: SD-9L/U-1, SD-11L/U-1, SD-12L/U-1, SD-13L/U-1, SD-13L/U-1, & SD-16L/U-1. Snare tube model numbers... on . The reason FDA records is "An investigation of consumer complaints revealed the inability of the snare wire to be retracted into the snare tube after deployment of the snare". As of , the recall status on record is Terminated.
Recall record
- Recall number
- Z-2236-2010
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- OLYMPUS Corporation of the Americas
- Product
- Olympus SD Snares, Reusable Electrosurgical Snares. Models: SD-9L/U-1, SD-11L/U-1, SD-12L/U-1, SD-13L/U-1, SD-13L/U-1, & SD-16L/U-1. Snare tube model numbers: MAJ-13 and MAJ-15. To be used with Olympus endoscopes to electrosurgically resect tissue within the digestive tract.
- Product code
- FDI Snare, Flexible
- Reason for recall
- An investigation of consumer complaints revealed the inability of the snare wire to be retracted into the snare tube after deployment of the snare.
- FDA root cause
- Other
- Action
- Firm initiated this recall following investigation of customer complaints regarding the inability of the snare wire to be retracted into the snare tube. Customers were notified via certified mail, an Urgent: Medical: Device Recall letter issued on May 4, 2010. The letter identified the affected product and the reason for the recall. Customers were instructed to inspect their inventory and identify the affected product. Customers are to remove from use any identified inventory, quarantine, and discard them. Customers should contact the firm to obtain replacement products. Questions should be directed towards Laura Storms-Tyler at 484-896-5688 or [email protected].
- Quantity in commerce
- 108 units
- Distribution
- Nationwide Distribution -- AR, CA, CO, CT, DC, FL, GA, IN, KS, LA, MD, MO, NC, NY, OH, PA, SC, TN, and TX.
- Product codes and lots
- Snare Package Lot numbers: 9YK, 9ZK, 01K, 02K, & 03K. Snare Tube Lot Numbers: K9Y10, K9Y16, K9Y18, K9Y19, KDY20, K9Y24, K9Y25, K9Y26, K9Y27, K9Y30,K9Z01, K9Z02, K9Z03, K9Z04, K9Z07, K9Z08, K9Z09, K9Z10, K9Z11, K9Z14,K9Z15, K9Z16, K9Z17, K9Z18, K9Z21, K9Z22, K9Z24, K9Z25, K9Z28, K9Z29, K0105, K0106, K0107, K0108, K0112, K0113, K0114, K0115, K0118, K0119, K0120, K0121, K0122, K0127, K0129, K0202, K0203, K0204, K0205, K0208, K0209, K0210, K0212, K0215, K0216, K0217, K0218, K0222, K0224,K0225, K0226, K0301, K0302, K0303, K0305, K0308, K0309, K0310, K0311, K0312, K0315, and K0316.