Z-1898-2010
Cepheid Xpert MRSA/SA Blood Culture Test for use with the GeneXpert Dx System, manufactured by Cepheid...
Cepheid / initiated 2010-04-05 / Terminated / root cause: manufacturing process
Cepheid began a recall of Cepheid Xpert MRSA/SA Blood Culture Test for use with the GeneXpert Dx System, manufactured by Cepheid, Sunnyvale, CA. Medical device, qualitative in vitro... on . The reason FDA records is "The firm is receiving reports of inaccurate results, which may result in delay of care for seriously ill patients". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-1898-2010
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Cepheid
- Product
- Cepheid Xpert MRSA/SA Blood Culture Test for use with the GeneXpert Dx System, manufactured by Cepheid, Sunnyvale, CA. Medical device, qualitative in vitro diagnostic test intended for the detection of Staphylococcus aureus and methicillin-resistant Staphylococcus aures DNA directly from patient positive blood cultures. It is not intended to monitor treatment for MRSA/SA infections. Sub culturing of positive blood cultures is necessary to recover organisms for susceptibility testing or for epidemiological typing.
- Product code
- NQX System, Nucleic Acid Amplification Test, Dna, Methicillin Resistant Staphylococcus Aureus, Direct Specimen
- Reason for recall
- The firm is receiving reports of inaccurate results, which may result in delay of care for seriously ill patients.
- Root cause tag
- manufacturing process
- FDA root cause
- Nonconforming Material/Component
- Action
- The firm, Cepheid, issued a technical bulletin to customers on April 5, 2010 and an updated to the April 5 bulletin dated April 9, 2010 instructing them to perform a reflexive repeat test on any blood culture results initially determined as having SA. If further information on this please contact your Cepheid Sales Representative 1-888-336-2743 or www.CEPHEID.COM.
- Quantity in commerce
- 5,510 units
- Distribution
- Worldwide distribution: USA and Europe, Taiwan, and Australia.
- Product codes and lots
- Product codes GXMRSA/SA-BC-10 (lot numbers 00601, 00602, 00603, 00701, 00801, 00901, 00902, 00903, 00904, 01001, 01101, 01102, 01301, 01302) and GXMRSA/SA-BC-CE-10 (lot numbers 01001, 01101, 01301)