Z-1898-2010

Cepheid Xpert MRSA/SA Blood Culture Test for use with the GeneXpert Dx System, manufactured by Cepheid...

Cepheid / initiated 2010-04-05 / Terminated / root cause: manufacturing process

Cepheid began a recall of Cepheid Xpert MRSA/SA Blood Culture Test for use with the GeneXpert Dx System, manufactured by Cepheid, Sunnyvale, CA. Medical device, qualitative in vitro... on . The reason FDA records is "The firm is receiving reports of inaccurate results, which may result in delay of care for seriously ill patients". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-1898-2010
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Cepheid
Product
Cepheid Xpert MRSA/SA Blood Culture Test for use with the GeneXpert Dx System, manufactured by Cepheid, Sunnyvale, CA. Medical device, qualitative in vitro diagnostic test intended for the detection of Staphylococcus aureus and methicillin-resistant Staphylococcus aures DNA directly from patient positive blood cultures. It is not intended to monitor treatment for MRSA/SA infections. Sub culturing of positive blood cultures is necessary to recover organisms for susceptibility testing or for epidemiological typing.
Product code
NQX System, Nucleic Acid Amplification Test, Dna, Methicillin Resistant Staphylococcus Aureus, Direct Specimen
Reason for recall
The firm is receiving reports of inaccurate results, which may result in delay of care for seriously ill patients.
Root cause tag
manufacturing process
FDA root cause
Nonconforming Material/Component
Action
The firm, Cepheid, issued a technical bulletin to customers on April 5, 2010 and an updated to the April 5 bulletin dated April 9, 2010 instructing them to perform a reflexive repeat test on any blood culture results initially determined as having SA. If further information on this please contact your Cepheid Sales Representative 1-888-336-2743 or www.CEPHEID.COM.
Quantity in commerce
5,510 units
Distribution
Worldwide distribution: USA and Europe, Taiwan, and Australia.
Product codes and lots
Product codes GXMRSA/SA-BC-10 (lot numbers 00601, 00602, 00603, 00701, 00801, 00901, 00902, 00903, 00904, 01001, 01101, 01102, 01301, 01302) and GXMRSA/SA-BC-CE-10 (lot numbers 01001, 01101, 01301)