Z-1900-2010

GENESIS II TIBIAL DRILL GUIDE, REF 71440238, SIZE 5, QTY: (1), NON-STERILE, Smith & Nephew, Inc., Memphis, TN...

Smith & Nephew, Inc. / initiated 2010-04-20 / Terminated / root cause: manufacturing process

Smith & Nephew Inc began a recall of GENESIS II TIBIAL DRILL GUIDE, REF 71440238, SIZE 5, QTY: (1), NON-STERILE, Smith & Nephew, Inc., Memphis, TN 38116, USA Drill guide used as manual surgical... on . The reason FDA records is "Markings for the drill guides are on the incorrect side: left is printed on right side and right is printed on the left side". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-1900-2010
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Smith & Nephew, Inc.
Product
GENESIS II TIBIAL DRILL GUIDE, REF 71440238, SIZE 5, QTY: (1), NON-STERILE, Smith & Nephew, Inc., Memphis, TN 38116, USA Drill guide used as manual surgical instrument during knee replacement surgeries.
Product code
LXH Orthopedic Manual Surgical Instrument
Reason for recall
Markings for the drill guides are on the incorrect side: left is printed on right side and right is printed on the left side
Root cause tag
manufacturing process
FDA root cause
Process control
Action
All affected Smith & Nephew consignees (International distributors) were notified of problem and the recall via email on 04/20/2010 and an Urgent: Medical Device Recall letter, dated 4/16/2010. The letter stated the reason for the recall and distributors were requested to immediately cease distribution and use of the affected products. They were also asked to examine their stock and stock of their accounts and quarantine any affected product. The Verification Section of the letter should be completed and returned even if there is no affected stock. Anyone the product was distributed to should also be notified and also return the product. Questions should be directed to Kelly Knight at 901-399-5441.
Quantity in commerce
12 units
Distribution
International Distribution Only -- Belgium, Spain, Germany, Australia, and Brazil.
Product codes and lots
Batch # 08JM07761 and 08JM07761A.