Z-0874-2012

Axium Detachable Coil System QC-7-30-Helix The ev3 Axium Detachable Coil System is indicated for the...

Medtronic, Inc. / initiated 2010-03-01 / Terminated / root cause: manufacturing process

Micro Therapeutics Inc, dba ev3 Neurovascular began a recall of Axium Detachable Coil System QC-7-30-Helix The ev3 Axium Detachable Coil System is indicated for the endovascular embolization of intracranial aneurysms and... on . The reason FDA records is "ev3 has initiated a voluntary removal of two lots of Axium Detachable Coil System for product mislabeling as the lots were swapped in production". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-0874-2012
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Medtronic, Inc.
Product
Axium Detachable Coil System QC-7-30-Helix The ev3 Axium Detachable Coil System is indicated for the endovascular embolization of intracranial aneurysms and other neurovascular abnormalities, such as arteriovenous malformations and arteriovenous fistulae.
Product code
KRD Device, Vascular, For Promoting Embolization
Reason for recall
ev3 has initiated a voluntary removal of two lots of Axium Detachable Coil System for product mislabeling as the lots were swapped in production.
Root cause tag
manufacturing process
FDA root cause
Employee error
Action
Micro Therapeutics Inc, dba ev3 Neurovascular performed product recovery according to the following outline. a) Quality Assurance Department (QAD) was used to provide details of product distribution for the affected lots. b) Based on the Health Hazard Assessment, the risk was considered undesirable. c) The affected product was part of a limited market release for Physician Preference Testing. As the product was within control of ev3 employees, no public notification was performed. d) Quality Systems contacted the sales representatives identifying the part number, lot number and number of units sent, and a request to segregate and return the affected product was initiated. e) Product returned to ev3 Neurovascular was held in Restricted Stock pending complete device accountability and disposition. f) The ev3 Neurovascular Quality Assurance department scrapped the affected product. g) The ev3 Neurovascular Regulatory Affairs department was responsible for assuring documentation and appropriate notification to regulatory bodies as required.
Quantity in commerce
14 units
Distribution
Worldwide Distribution - USA (nationwide) and the countries of the UK, and France.
Product codes and lots
Lot # 7964590