Z-1806-2010
Calcium Sulfate Hemihydrate 1.0 gram kits
Ace Surgical Supply Co., Inc. / initiated 2010-05-14 / Terminated / root cause: manufacturing process
Ace Surgical Supply Co., Inc. began a recall of Calcium Sulfate Hemihydrate 1.0 gram kits on . The reason FDA records is "Fast set 4% potassium sulfate solution may not be sterile. For customers who have used the Calcium Sulfate Hemihydrate kit without the Fast Set solution, then these potential risks identified do not apply. Risks of...". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-1806-2010
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Ace Surgical Supply Co., Inc.
- Product
- Calcium Sulfate Hemihydrate 1.0 gram kits.
- Product code
- LYC Bone Grafting Material, Synthetic
- Reason for recall
- Fast set 4% potassium sulfate solution may not be sterile. For customers who have used the Calcium Sulfate Hemihydrate kit without the Fast Set solution, then these potential risks identified do not apply. Risks of adverse health consequences with the Fast Set 4% potassium sulfate solution include tissue inflammation and infection.
- Root cause tag
- manufacturing process
- FDA root cause
- Process control
- Action
- Ace Surgical has become aware through internal review of possibility sterility issues with one of the components included in the Calcium Hemihydrate 0.5 and 1.0 gram kits. Customers with questions should contact ACE Surgical using the recall hot line at (800) 441-3100 Ext. 204.
- Quantity in commerce
- 7930
- Distribution
- Class 2 recall - Worldwide distribution . Item Number 505-2001: South Africa, Australia, Vietnam, Ontario, Canada, Italy, Lebanon, Petaling Jaya, Malaysia, Portugal, and West Indies. Item Number 505-2002: Australia, Bermuda, Canada, Lebanon, Petaling Jaya, Malaysia, and Portugal.
- Product codes and lots
- Calcium Sulfate Hemihydrate 1.0 gram Product Code #5052001 lot numbers: 08030106, 08030030, 08040035, 08050025, 08070036, 08090049, 08100092, 08110041, 08120097, 09010055, 09020041, 09030047, 09040017, 09050004, 09060013, 09060095, 09070046, 09080033, 09090026, 09100002, 09100066, 09120011, 09120070, 10010007, 10010067, 10020064, 10030003, 10030060, 10040011, 10040046