Z-2398-2010
Clinical Chemistry Phosphorus assay
Abbott Medical / initiated 2010-05-04 / Terminated / root cause: manufacturing process
Abbott Laboratories Inc. began a recall of Clinical Chemistry Phosphorus assay on . The reason FDA records is "Product may give inaccurate results and aspiration errors with reagent configuration, including results outside of linear range & calibration failure". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-2398-2010
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Abbott Medical
- Product
- Clinical Chemistry Phosphorus assay
- Product code
- CEO Phosphomolybdate (Colorimetric), Inorganic Phosphorus
- Reason for recall
- Product may give inaccurate results and aspiration errors with reagent configuration, including results outside of linear range & calibration failure.
- Root cause tag
- manufacturing process
- FDA root cause
- Other
- Action
- Abbott Laboratories issued a "Product Recall: Immediate Action Required" letter dated May 4, 2010 to consignees. The letter described the product, problem and action to be taken by the customer. The letter instructed the customer to: 1) Identify affected product in inventory 2) Discontinue use and destroy any remaining inventory of the affected lot: 7D71-21 3) Complete and return an enclosed Customer Reply Form 4) Order alternative kit sizes for Clinical Chemistry Phosphorus: LN 7D71-31 or 7D71-22 5) For AEROSET Systems, manually update parameters as described in the letter. 6) For ARCHITECT cSystems, implement alternative kit sizes without system changes. 7) Provide a copy of the letter to other laboratories that have been distributed the affected product. US customers can contact Abbott Customer Service at 1-877-4ABBOTT. Customers outside the U>S> are instructed to contact their local area Customer Service.
- Quantity in commerce
- 3136
- Distribution
- Worldwide Distribution: US, including Puerto Rico, and countries of Germany, New Zealand, Trinidad/Tobago, Cayman Island, Hong Kong, Chile, Curacao, Jamaica, Australia, and Bahamas.
- Product codes and lots
- Lot number: 81040HW00; expires July 15, 2010, 85009HW00; expires December 15th, 2010. (All Lots of Clincal Phosphorous LN 7D71-21).