Z-2398-2010

Clinical Chemistry Phosphorus assay

Abbott Medical / initiated 2010-05-04 / Terminated / root cause: manufacturing process

Abbott Laboratories Inc. began a recall of Clinical Chemistry Phosphorus assay on . The reason FDA records is "Product may give inaccurate results and aspiration errors with reagent configuration, including results outside of linear range & calibration failure". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-2398-2010
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Abbott Medical
Product
Clinical Chemistry Phosphorus assay
Product code
CEO Phosphomolybdate (Colorimetric), Inorganic Phosphorus
Reason for recall
Product may give inaccurate results and aspiration errors with reagent configuration, including results outside of linear range & calibration failure.
Root cause tag
manufacturing process
FDA root cause
Other
Action
Abbott Laboratories issued a "Product Recall: Immediate Action Required" letter dated May 4, 2010 to consignees. The letter described the product, problem and action to be taken by the customer. The letter instructed the customer to: 1) Identify affected product in inventory 2) Discontinue use and destroy any remaining inventory of the affected lot: 7D71-21 3) Complete and return an enclosed Customer Reply Form 4) Order alternative kit sizes for Clinical Chemistry Phosphorus: LN 7D71-31 or 7D71-22 5) For AEROSET Systems, manually update parameters as described in the letter. 6) For ARCHITECT cSystems, implement alternative kit sizes without system changes. 7) Provide a copy of the letter to other laboratories that have been distributed the affected product. US customers can contact Abbott Customer Service at 1-877-4ABBOTT. Customers outside the U>S> are instructed to contact their local area Customer Service.
Quantity in commerce
3136
Distribution
Worldwide Distribution: US, including Puerto Rico, and countries of Germany, New Zealand, Trinidad/Tobago, Cayman Island, Hong Kong, Chile, Curacao, Jamaica, Australia, and Bahamas.
Product codes and lots
Lot number: 81040HW00; expires July 15, 2010, 85009HW00; expires December 15th, 2010. (All Lots of Clincal Phosphorous LN 7D71-21).