ANGIODYNAMICS BENEPHIT(TM) XT Infusion System, 35 cm, Targeted Renal Therapy Infusion system, Model/REF...
Angiodynamics, Inc. began a recall of ANGIODYNAMICS BENEPHIT(TM) XT Infusion System, 35 cm, Targeted Renal Therapy Infusion system, Model/REF Number 70035, STERILE, EO, Intended for Single Use... on . The reason FDA records is "Benephit Infusion Catheters may exhibit a hole or tear in the primary sterile barrier packaging. The product is intended to be sterile. Breaches in the sterile packaging may result in non-sterile product, which could...". As of , the recall status on record is Terminated, and MDM tags the root cause as packaging.
Recall record
- Recall number
- Z-1733-2010
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Boston Scientific Corporation
- Product
- ANGIODYNAMICS BENEPHIT(TM) XT Infusion System, 35 cm, Targeted Renal Therapy Infusion system, Model/REF Number 70035, STERILE, EO, Intended for Single Use Only. AngioDynamics, Incorporated 603 Queensbury, NY 12804.
- Product code
- KRA Catheter, Continuous Flush
- Reason for recall
- Benephit Infusion Catheters may exhibit a hole or tear in the primary sterile barrier packaging. The product is intended to be sterile. Breaches in the sterile packaging may result in non-sterile product, which could potentially cause a patient infection which is a serious medical condition.
- Root cause tag
- packaging
- FDA root cause
- Packaging change control
- Action
- Angiodynamics, Inc. issued an "Urgent Medical Device Recall" notification (dated 4/28/10). Customers were advised of the affected product and instructed to return all recalled product to the firm. For further information, contact Angiodynamics, Inc. Customer Service at 1-800-772-6446.
- Quantity in commerce
- 741
- Distribution
- Distributed Nationwide and to Italy
- Product codes and lots
- Batch Numbers: P909081, P909291, P910092, P910262, P911053, P911163, P911243, P912083, P912293 and P001203.