Z-1940-2010

BD Biosciences, BD FACS Sample Prep Assistant II (SPA II), blood sample preparer for flow cytometry analysis

Becton, Dickinson and Company, BD Biosciences / initiated 2010-02-19 / Terminated / root cause: manufacturing process

BD Biosciences, Systems & Reagents began a recall of BD Biosciences, BD FACS Sample Prep Assistant II (SPA II), blood sample preparer for flow cytometry analysis on . The reason FDA records is "Samples prepared on the BD SPA II, SN T0076 produced inaccurate sample results due to inaccurate sample dispensed as a result of the cap piercing probe". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-1940-2010
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Becton, Dickinson and Company, BD Biosciences
Product
BD Biosciences, BD FACS Sample Prep Assistant II (SPA II), blood sample preparer for flow cytometry analysis
Product code
GKZ Counter, Differential Cell
Reason for recall
Samples prepared on the BD SPA II, SN T0076 produced inaccurate sample results due to inaccurate sample dispensed as a result of the cap piercing probe
Root cause tag
manufacturing process
FDA root cause
Nonconforming Material/Component
Action
The firm filed an MDR and implemented a field action consisting of a February 2010 notification letter and product replacement of affected probes. A CAPA for the probe design has been implemented. Effectivity is being tracked via mail and follow up phone call.
Quantity in commerce
349 units
Distribution
Worldwide distribution: USA, Australia, Belgium, China (Hong Kong), Colombia, Jamaica, Korea, Mexico, Singapore and Uruguay.
Product codes and lots
Sample Probe Catalog Numbers 337170, 333494, 333485, 339118, 647768, 647769, and lot numbers , 92035 and 92036.