Z-1773-2010
Boston Scientific Radial Jaw 4 Large Capacity w/Needle Biopsy Forceps Box 80, 240cm UPN/Catalog Number...
Boston Scientific Corporation / initiated 2010-05-06 / Terminated / root cause: manufacturing process
Boston Scientific Corporation began a recall of Boston Scientific Radial Jaw 4 Large Capacity w/Needle Biopsy Forceps Box 80, 240cm UPN/Catalog Number: M00513334 on . The reason FDA records is "Clevis may detach from the coil preventing the device to open and/or close. Delays in the procedure may result". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-1773-2010
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Boston Scientific Corporation
- Product
- Boston Scientific Radial Jaw 4 Large Capacity w/Needle Biopsy Forceps Box 80, 240cm UPN/Catalog Number: M00513334
- Product code
- FCL Forceps, Biopsy, Non-Electric
- Reason for recall
- Clevis may detach from the coil preventing the device to open and/or close. Delays in the procedure may result.
- Root cause tag
- manufacturing process
- FDA root cause
- Process control
- Action
- Boston Scientific issued Recall notification letters via Federal Express Priority mail on May 6, 2010. The notification instructs the account to check their inventory, remove any affected product from their inventory, and return it to Boston Scientific. The notification also requests the completion of a Reply Verification Tracking Form as a method of documenting the presence or absence of affected product from their inventory. In the letter, customers with questions are provided with a phone and email at the firm to contact.
- Distribution
- Nationwide; Panama.
- Product codes and lots
- Lots: 13361332, 13361334