Z-1969-2010

Starion Instruments TLS3 35C Thermal Ligating Shears, Electrosurgical cutting and coagulation device and...

Microline Surgical, Inc / initiated 2008-02-06 / Terminated / root cause: manufacturing process

Starion Instruments began a recall of Starion Instruments TLS3 35C Thermal Ligating Shears, Electrosurgical cutting and coagulation device and accessories, manufactured by Starion Instruments Corp... on . The reason FDA records is "Some jaws exhibited lack of ductility and could break if overloaded or used improperly by placing too much tissue in the bottom of the "v" of the jaw". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-1969-2010
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Microline Surgical, Inc
Product
Starion Instruments TLS3 35C Thermal Ligating Shears, Electrosurgical cutting and coagulation device and accessories, manufactured by Starion Instruments Corp, Sunnyvale, CA. Intended use: medical device for use in cutting and cauterization of soft tissue during surgery, as well as cutting of natural or synthetic, non-metallic sutures during surgery.
Product code
GEI Electrosurgical, Cutting & Coagulation & Accessories
Reason for recall
Some jaws exhibited lack of ductility and could break if overloaded or used improperly by placing too much tissue in the bottom of the "v" of the jaw.
Root cause tag
manufacturing process
FDA root cause
Nonconforming Material/Component
Action
The firm called domestic consignees but no letter was sent to domestic cosignees and a notification letter was sent to international consignees. The firm called domestic consignees and sent a notification letter to international consignees. No records were kept of phone calls and a copy of the letter sent to international consignees was not obtained
Distribution
Worldwide distribution
Product codes and lots
Model number 132-136D, lot numbers 708027 and 710008, expiration dates 09/21/12 and 10/19/12