Starion Instruments Real Hand ThermaSeal Thermal Ligating Shears, electrosurgical cutting and coagulation...
Starion Instruments began a recall of Starion Instruments Real Hand ThermaSeal Thermal Ligating Shears, electrosurgical cutting and coagulation device and accessories, manufactured by Starion... on . The reason FDA records is "Due to an improper fixture revision, device may be missing a screw at the jaw assembly, resulting in jaw components falling off during use". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-1972-2010
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Microline Surgical, Inc
- Product
- Starion Instruments Real Hand ThermaSeal Thermal Ligating Shears, electrosurgical cutting and coagulation device and accessories, manufactured by Starion Instruments, Sunnyvale, CA Intended use: Medical Device used for cutting and cauterization of soft tissue during surgery, and of natural or synthetic, non-metallic sutures during surgery.
- Product code
- GEI Electrosurgical, Cutting & Coagulation & Accessories
- Reason for recall
- Due to an improper fixture revision, device may be missing a screw at the jaw assembly, resulting in jaw components falling off during use.
- Root cause tag
- manufacturing process
- FDA root cause
- Process control
- Action
- The firm submitted an MDR but determined that there was no risk to health. E-mails were sent on June 26, 2008, to consignees, describing the product, the problem, and action to be taken by consignees. The consignees were ask to provide a list of all unused ThermaSeal devices and their locations, to return the products to Starion from the lots noted and to provide a copy of the attached letter to their sales force and customers if they choose. If you have any questions, please call (408) 522-5200 ext. 326 or (408) 350-9934.
- Quantity in commerce
- 217 units
- Distribution
- Worldwide distribution.
- Product codes and lots
- Model Number N-10700, lot numbers 80419, 806009, 804037, 804035, 804036, 804019A.