Z-2423-2010

Siemens Healthcare Diagnostics, Inc., Dimension Vista Intelligent Laboratory System with Alpha Pinnacle 10kVA...

Siemens Healthcare Diagnostics Inc. / initiated 2010-05-05 / Terminated / root cause: manufacturing process

Siemens Healthcare Diagnostics, Inc. began a recall of Siemens Healthcare Diagnostics, Inc., Dimension Vista Intelligent Laboratory System with Alpha Pinnacle 10kVA UPS External Battery Pack. Part numbers... on . The reason FDA records is "Defective batteries used in the Alpha Pinnacle P10 UPS External Battery Pack may overheat and cause a strong sulfuric odor to be released and a corrosive sulfuric acid solution leak to occur". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-2423-2010
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Siemens Healthcare Diagnostics Inc.
Product
Siemens Healthcare Diagnostics, Inc., Dimension Vista Intelligent Laboratory System with Alpha Pinnacle 10kVA UPS External Battery Pack. Part numbers: 070200001, 070200011, 070200021; Dimension Vista 1500 and 3000T instruments.
Product code
JJE Analyzer, Chemistry (Photometric, Discrete), For Clinical Use
Reason for recall
Defective batteries used in the Alpha Pinnacle P10 UPS External Battery Pack may overheat and cause a strong sulfuric odor to be released and a corrosive sulfuric acid solution leak to occur.
Root cause tag
manufacturing process
FDA root cause
Nonconforming Material/Component
Action
Siemens issued an Urgent Field Safety Notice in May 2010 to customers with information on what to do if the battery leaks, overheats, or vents an odor. Firm also issued a Support Bulletin to field service personnel to inspect and replace affected units.
Quantity in commerce
228 units
Distribution
Worldwide distribution: USA, Australia, Austria, Belgium, Canada, France, Germany, Italy, Japan, Malaysia, Netherlands, New Zealand, Portugal, Slovenia, Spain, and Switzerland.
Product codes and lots
Part numbers: 070200001, 070200011, 070200021; Dimension Vista 1500 and 3000T instruments within the serial number range of B01060900055 through B01090100414R0, excluding B01081100388R0, B01081100389R0, B01081200396R0, B01081200402R0, B01081200403R0, B01081200404R0, B01081200406R0, B01081200408R0, B01081200410R0, B01081200411R0, B01090100412R0, B01090100413R0.