Z-2288-2010
ReliaMed Deluxe Portable Adjusting Table. Product is a medical table used on patients for massage and...
MDC Acquisition Co. / initiated 2010-04-23 / Terminated / root cause: design
MDC Acquisition Co. began a recall of ReliaMed Deluxe Portable Adjusting Table. Product is a medical table used on patients for massage and chiropractic treatment on . The reason FDA records is "Two complaints were received of the vinyl-covered, padded, plywood table surface breaking near the foot end". As of , the recall status on record is Terminated, and MDM tags the root cause as design.
Recall record
- Recall number
- Z-2288-2010
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- MDC Acquisition Co.
- Product
- ReliaMed Deluxe Portable Adjusting Table. Product is a medical table used on patients for massage and chiropractic treatment.
- Product code
- FPS Board, Bed
- Reason for recall
- Two complaints were received of the vinyl-covered, padded, plywood table surface breaking near the foot end.
- Root cause tag
- design
- FDA root cause
- Component design/selection
- Action
- "Urgent: Medical Device Recall" letters were sent to consignees from Milliken Medical and Meyer Distributing Company in April 2010 describing the issue. They instructed their customers to immediately stop use of the table. customers were instructed to return the tables to the firm with an enclosed Recall Form for future replacement, or for replacement with a comparable table from another menufacturer at no cost, or for refund of the original cost of the device. Consignees may contact Meyer Distributing Company at 800-472-4221 and Milliken Medical at 800-532-1356.
- Quantity in commerce
- 265 Tables
- Distribution
- Worldwide distribution: USA, including the states of: AK, AL, AR, AZ, CA, CO, CT, DC, FL, GA, HI, IA, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, NH, NJ, NM, NV, NY, OH, OK, OR, PA, SC, TN, TX, VA, WA and WI, and the country of Canada.
- Product codes and lots
- The table do not bear lot or serial numbers; however there are two device identification numbers representative of two different colors. ZZRNT100BL (blue) ZZRMT100BLK (black)