Artificial ligament fixation device # 7PE ZIPLOOP Extended Toggle, Sterile, REF 904755, Biomet Sports...
Biomet, Inc. began a recall of Artificial ligament fixation device # 7PE ZIPLOOP Extended Toggle, Sterile, REF 904755, Biomet Sports Medicine Warsaw, IN. Intended for soft tissue to bone... on . The reason FDA records is "The firm has received complaints that the saddle is too long. The saddle is supposed to be between 7/8 and 1 1/8 inches long. The saddle in the complaints was up to 1.9 inches long". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-1952-2010
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Zimmer GmbH
- Product
- Artificial ligament fixation device # 7PE ZIPLOOP Extended Toggle, Sterile, REF 904755, Biomet Sports Medicine Warsaw, IN. Intended for soft tissue to bone fixation.
- Product code
- MBI Fastener, Fixation, Nondegradable, Soft Tissue
- Reason for recall
- The firm has received complaints that the saddle is too long. The saddle is supposed to be between 7/8 and 1 1/8 inches long. The saddle in the complaints was up to 1.9 inches long.
- Root cause tag
- manufacturing process
- FDA root cause
- Process control
- Action
- Biomet Sports Medicine issued an "Urgent Medical Device Recall Notice" dated March 31, 2010 to user facilities and distributors. A second notification containing updated information was issued June 2, 2010. Consignees were instructed to discontinue use of the product and return it to the firm along with a completed Response Form. For further information, contact Biomet at 1-574-371-3755 or 1-574-371-3756, Monday through Friday, 8AM to 5PM.
- Quantity in commerce
- 2466 both products
- Distribution
- Worldwide Distribution -- United States, Canada, Australia, The Netherlands, Belgium, Italy, France, New Zealand, Austria, United Kingdom, Chile, Greece, Poland, Denmark, Turkey, Portugal, South Africa and Spain.
- Product codes and lots
- Lot Numbers: 047510, 050400, 050420, 089710, 089800, 182820, 192240, 192250, 192280, 219260, 219270, 219280, 228530, 228550, 228580, 228590, 228610, 228620, 295450, 361020, 361020, 392930, 392990, 483490, 483500, 529380, 527780, 529800, 529810, 582890, 582910, 627510, 648470, 648480, 671130, 671150, 671170, 671180, 671190, 671240, 706150, 706160, 782360, 782370, 821710, 821730, 943000, 943010, 943030, 952040, 952070 and 974140.