Toggleloc Artificial ligament fixation device ZIPLOOP Technology Implant Kit, Sterile, REF 909848, Biomet...
Biomet, Inc. began a recall of Toggleloc Artificial ligament fixation device ZIPLOOP Technology Implant Kit, Sterile, REF 909848, Biomet Sports Medicine Warsaw, IN. Intended for soft tissue... on . The reason FDA records is "The firm has received complaints that the saddle is too long. The saddle is supposed to be between 7/8 and 1 1/8 inches long. The saddle in the complaints was up to 1.9 inches long". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-1953-2010
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Zimmer GmbH
- Product
- Toggleloc Artificial ligament fixation device ZIPLOOP Technology Implant Kit, Sterile, REF 909848, Biomet Sports Medicine Warsaw, IN. Intended for soft tissue to bone fixation.
- Product code
- MBI Fastener, Fixation, Nondegradable, Soft Tissue
- Reason for recall
- The firm has received complaints that the saddle is too long. The saddle is supposed to be between 7/8 and 1 1/8 inches long. The saddle in the complaints was up to 1.9 inches long.
- Root cause tag
- manufacturing process
- FDA root cause
- Process control
- Action
- Biomet Sports Medicine issued an "Urgent Medical Device Recall Notice" dated March 31, 2010 to user facilities and distributors. A second notification containing updated information was issued June 2, 2010. Consignees were instructed to discontinue use of the product and return it to the firm along with a completed Response Form. For further information, contact Biomet at 1-574-371-3755 or 1-574-371-3756, Monday through Friday, 8AM to 5PM.
- Quantity in commerce
- 2466 both products
- Distribution
- Worldwide Distribution -- United States, Canada, Australia, The Netherlands, Belgium, Italy, France, New Zealand, Austria, United Kingdom, Chile, Greece, Poland, Denmark, Turkey, Portugal, South Africa and Spain.
- Product codes and lots
- Lot Numbers: 228150, 228160, 783930, 975220, 975230 and 975240.