IMMAGE Immunochemistry Systems Buffer 1 (BUF1), Part Number: 447650
Beckman Coulter Inc. began a recall of IMMAGE Immunochemistry Systems Buffer 1 (BUF1), Part Number: 447650 on . The reason FDA records is "Beckman Coulter is initiating a recall of IMMAGE Immunochemistry Systems Buffer 1 (BUF1) due to received reports of increased recovery of control and patient samples using the lots of IMMAGE BUF1. Some lots of BUF1 have...". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-2195-2010
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Siemens Healthcare Diagnostics Inc.
- Product
- IMMAGE Immunochemistry Systems Buffer 1 (BUF1), Part Number: 447650
- Product code
- JQX Nephelometer, For Clinical Use
- Reason for recall
- Beckman Coulter is initiating a recall of IMMAGE Immunochemistry Systems Buffer 1 (BUF1) due to received reports of increased recovery of control and patient samples using the lots of IMMAGE BUF1. Some lots of BUF1 have been reported to cause high shifts in recovery of control or patient samples. Quality Control failures may result after replacing BUF1 with a different lot.
- Root cause tag
- manufacturing process
- FDA root cause
- Nonconforming Material/Component
- Action
- Beckman Coulter will send a Product Corrective Action (PCA) letter on the week of April 19, 2010 to the affected customers. The letter provided the customers with an explanation of the problem and a work around.
- Quantity in commerce
- 1309
- Distribution
- Worldwide: U.S., Canada, Australia, Austria, Belgium, China, Czech Republic, Denmark, France, Germany, Greece, Hong Kong, Iceland, Ireland, Italy, Japan, Liechtenstein, Macao, Malaysia, Netherlands, New Zealand, Norway, Oman, Poland, Qatar, Russian Federation, Singapore, Slovakia, South Africa, Spain, Sweden, Switzerland, Taiwan, Tunisia, Turkey and United Kingdom,
- Product codes and lots
- Lot#s: T908069, T908104, T909110, T001036 ( This lot was not shipped to customers.)