Z-2195-2010

IMMAGE Immunochemistry Systems Buffer 1 (BUF1), Part Number: 447650

Siemens Healthcare Diagnostics Inc. / initiated 2010-05-21 / Terminated / root cause: manufacturing process

Beckman Coulter Inc. began a recall of IMMAGE Immunochemistry Systems Buffer 1 (BUF1), Part Number: 447650 on . The reason FDA records is "Beckman Coulter is initiating a recall of IMMAGE Immunochemistry Systems Buffer 1 (BUF1) due to received reports of increased recovery of control and patient samples using the lots of IMMAGE BUF1. Some lots of BUF1 have...". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-2195-2010
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Siemens Healthcare Diagnostics Inc.
Product
IMMAGE Immunochemistry Systems Buffer 1 (BUF1), Part Number: 447650
Product code
JQX Nephelometer, For Clinical Use
Reason for recall
Beckman Coulter is initiating a recall of IMMAGE Immunochemistry Systems Buffer 1 (BUF1) due to received reports of increased recovery of control and patient samples using the lots of IMMAGE BUF1. Some lots of BUF1 have been reported to cause high shifts in recovery of control or patient samples. Quality Control failures may result after replacing BUF1 with a different lot.
Root cause tag
manufacturing process
FDA root cause
Nonconforming Material/Component
Action
Beckman Coulter will send a Product Corrective Action (PCA) letter on the week of April 19, 2010 to the affected customers. The letter provided the customers with an explanation of the problem and a work around.
Quantity in commerce
1309
Distribution
Worldwide: U.S., Canada, Australia, Austria, Belgium, China, Czech Republic, Denmark, France, Germany, Greece, Hong Kong, Iceland, Ireland, Italy, Japan, Liechtenstein, Macao, Malaysia, Netherlands, New Zealand, Norway, Oman, Poland, Qatar, Russian Federation, Singapore, Slovakia, South Africa, Spain, Sweden, Switzerland, Taiwan, Tunisia, Turkey and United Kingdom,
Product codes and lots
Lot#s: T908069, T908104, T909110, T001036 ( This lot was not shipped to customers.)