Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-2009-2010

Microstream SCP H PLUS O2 ADU/INT 100UN Part Number: 010625

Oridion Medical 1987 Ltd. / initiated 2010-05-26 / Terminated / root cause: manufacturing process

Oridion Medical 1987 Ltd. began a recall of Microstream SCP H PLUS O2 ADU/INT 100UN Part Number: 010625 on . The reason FDA records is "The Microstream Filterline may not be recognized by the monitor when connected. No CO2 readings will be displayed and no alarms may sound for unmonitored patients. Replacing the filterline may delay treatment". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-2009-2010
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Oridion Medical 1987 Ltd.
Product
Microstream SCP H PLUS O2 ADU/INT 100UN Part Number: 010625
Product code
cck
Reason for recall
The Microstream Filterline may not be recognized by the monitor when connected. No CO2 readings will be displayed and no alarms may sound for unmonitored patients. Replacing the filterline may delay treatment.
Root cause tag
manufacturing process
FDA root cause
Process design
Action
Oridion notified customers by letter on 5/26/10 and requested that they remove the FilterLines from the related production lots from the field and an immediate replacement with FilterLines from subsequent production lots. Questions are directed to the responsible person at the supplier named in the letter, or to Oridion Capnography at 888-674-3466.
Quantity in commerce
13500 units (13 boxes/100 units)
Distribution
AZ, CA, FL, HI, LA, , MA, MO, NJ, OH, PA, WA, WI,
Product codes and lots
Lot numbers from Q0901006 to Q0908223. Produced between January and July 2009,