Microstream FILTERLINE SET ADULT/PEDIATRIC 100 UN Part Number: 010579-01
Oridion Medical 1987 Ltd. began a recall of Microstream FILTERLINE SET ADULT/PEDIATRIC 100 UN Part Number: 010579-01 on . The reason FDA records is "The Microstream Filterline may not be recognized by the monitor when connected. No CO2 readings will be displayed and no alarms may sound for unmonitored patients. Replacing the filterline may delay treatment". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-2010-2010
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Oridion Medical 1987 Ltd.
- Product
- Microstream FILTERLINE SET ADULT/PEDIATRIC 100 UN Part Number: 010579-01
- Product code
- cck
- Reason for recall
- The Microstream Filterline may not be recognized by the monitor when connected. No CO2 readings will be displayed and no alarms may sound for unmonitored patients. Replacing the filterline may delay treatment.
- Root cause tag
- manufacturing process
- FDA root cause
- Process design
- Action
- Oridion notified customers by letter on 5/26/10 and requested that they remove the FilterLines from the related production lots from the field and an immediate replacement with FilterLines from subsequent production lots. Questions are directed to the responsible person at the supplier named in the letter, or to Oridion Capnography at 888-674-3466.
- Quantity in commerce
- 1300 units (13 boxes/100 units)
- Distribution
- AZ, CA, FL, HI, LA, , MA, MO, NJ, OH, PA, WA, WI,
- Product codes and lots
- Lot numbers from Q0901006 to Q0908223. Produced between January and July 2009,