Cepheid Xpert Flu A Panel (GXFLUPANEL-CE-10), manufactured by Cepheid, Sunnyvale, CA. Indicated as an...
Cepheid began a recall of Cepheid Xpert Flu A Panel (GXFLUPANEL-CE-10), manufactured by Cepheid, Sunnyvale, CA. Indicated as an automated multiplex real-time RT-PCR assay for use in... on . The reason FDA records is "Firm has had an increasing number of customer complaints regarding higher than expected Indeterminate Test Results due to "Invalid" test reports from our system when they have been conducting their own internal...". As of , the recall status on record is Terminated.
Recall record
- Recall number
- Z-1652-2010
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Cepheid
- Product
- Cepheid Xpert Flu A Panel (GXFLUPANEL-CE-10), manufactured by Cepheid, Sunnyvale, CA. Indicated as an automated multiplex real-time RT-PCR assay for use in labroatories for the in vitro qualitative detection and differentiation of 2009 H1N1 influenza viral RNA.
- Product code
- OPU
- Reason for recall
- Firm has had an increasing number of customer complaints regarding higher than expected Indeterminate Test Results due to "Invalid" test reports from our system when they have been conducting their own internal validation studies using banked frozen repository specimens.
- FDA root cause
- Other
- Action
- Cepheid issued a notification to consignees regarding the affected product. Additional instructions were provided to consignees in order to avoid further "Indeterminate Results". For further information, contact Cepheid at 1-888-336-2743.
- Quantity in commerce
- 944 kits
- Distribution
- Worldwide Distribution -- United States, France, Belgium, Germany, Norway Austria, Denmark, Spain, Taiwan and Hong Kong.
- Product codes and lots
- Lot Number: 00501 (Expires 2010 08 08)