Z-2223-2010

Disposable Surgical Cable

Remington Medical / initiated 2010-05-19 / Terminated / root cause: manufacturing process

Remington Medical Inc. began a recall of Disposable Surgical Cable, Product Code: FL-601-97 (100), Single use, Rx Only, Remington Medical, Inc., 6830 Meadowridge Court, Alpharetta, GA 30005 Usage... on . The reason FDA records is "The Safe Connect component may separate from the cable, causing the cable to become unplugged from the EPG". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-2223-2010
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Remington Medical
Product
Disposable Surgical Cable, Product Code: FL-601-97 (100), Single use, Rx Only, Remington Medical, Inc., 6830 Meadowridge Court, Alpharetta, GA 30005 Usage: Temporary pacing of open hearts: used with External Pacing Generators (EPGs).
Product code
DSA Cable, Transducer And Electrode, Patient, (Including Connector)
Reason for recall
The Safe Connect component may separate from the cable, causing the cable to become unplugged from the EPG.
Root cause tag
manufacturing process
FDA root cause
Process design
Action
Consignees were notified on/about 05/19/2010. They were instructed to examine their inventory and quarantine product subject to the recall. If they distributed the product, they were advised to identify their customers and notify them of the recall (via the recall letter). A Return Authorization was provided. More information can be obtained by calling 770.888.8520.
Quantity in commerce
1488 units
Distribution
AZ, CA, CO, CT, DE, FL, GA, IL, IN, LA, MD, MI, MO, MS, NC, NM, NY, OH, OK, OR, PA, SC, TX, TN, TX, VA, WA, WI and Canada.
Product codes and lots
Lot number: 100701