Z-1839-2010
Ultraflex Tracheobronchial Stent System. Uncovered - 20/80 Catalog Number: M00569310. Single Use, Rx only...
Boston Scientific Corporation / initiated 2010-05-17 / Terminated / root cause: design
Boston Scientific Corporation began a recall of Ultraflex Tracheobronchial Stent System. Uncovered - 20/80 Catalog Number: M00569310. Single Use, Rx only. Boston Scientific, Natick, MA 01760. Indicated for... on . The reason FDA records is "Suture related device failures during deployment and repositioning of the stent". As of , the recall status on record is Terminated, and MDM tags the root cause as design.
Recall record
- Recall number
- Z-1839-2010
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Boston Scientific Corporation
- Product
- Ultraflex Tracheobronchial Stent System. Uncovered - 20/80 Catalog Number: M00569310. Single Use, Rx only. Boston Scientific, Natick, MA 01760. Indicated for the use in the treatment of tracheobronchial structures produced by malignant neoplasms.
- Product code
- JCT Prosthesis, Tracheal, Expandable
- Reason for recall
- Suture related device failures during deployment and repositioning of the stent
- Root cause tag
- design
- FDA root cause
- Device Design
- Action
- Boston Scientific issued an "Urgent Medical Device Recall - Immediate Action Required" notification dated May 17, 2010 to U.S. Distributors stating that use and distribution of products should cease immediately and return all affected product to the firm for replacement. Distributors were asked to complete and return reply a verification tracking form. For further information, contact Boston Scientific at 1-866-868-4004.
- Distribution
- Worldwide Distribution - Only US distributed product is affected by recall
- Product codes and lots
- Lot Numbers: 9820027 9820415 11092859 11093048 11752551 11962126 11962127