Z-1840-2010

Ultraflex Tracheobronchial Stent System. Covered - 8/40/25 Catalog Number: M00569410. Single Use, Rx only...

Boston Scientific Corporation / initiated 2010-05-17 / Terminated / root cause: design

Boston Scientific Corporation began a recall of Ultraflex Tracheobronchial Stent System. Covered - 8/40/25 Catalog Number: M00569410. Single Use, Rx only. Boston Scientific, Natick, MA 01760. Indicated for... on . The reason FDA records is "Suture related device failures during deployment and repositioning of the stent". As of , the recall status on record is Terminated, and MDM tags the root cause as design.

Recall record

Recall number
Z-1840-2010
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Boston Scientific Corporation
Product
Ultraflex Tracheobronchial Stent System. Covered - 8/40/25 Catalog Number: M00569410. Single Use, Rx only. Boston Scientific, Natick, MA 01760. Indicated for the use in the treatment of tracheobronchial structures produced by malignant neoplasms.
Product code
JCT Prosthesis, Tracheal, Expandable
Reason for recall
Suture related device failures during deployment and repositioning of the stent
Root cause tag
design
FDA root cause
Device Design
Action
Boston Scientific issued an "Urgent Medical Device Recall - Immediate Action Required" notification dated May 17, 2010 to U.S. Distributors stating that use and distribution of products should cease immediately and return all affected product to the firm for replacement. Distributors were asked to complete and return reply a verification tracking form. For further information, contact Boston Scientific at 1-866-868-4004.
Distribution
Worldwide Distribution - Only US distributed product is affected by recall
Product codes and lots
Lot Numbers: 9719739 11024662 11026969 11141579 11142082 11234436 11234516 11579871 11580155 11629573 11720857 11720859 11770928 11770932 11805568 11805570 11963314 11963318 12160054 12160056 12316137 12371714 12403737 12464577 12588801 12639049 12853698 12906234 13073202 13104729 13200936