Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-1860-2010

Ultraflex Esophageal Stent System. Uncovered - Distal Release - 18/23/15. Catalog Number M00513720. Single...

Boston Scientific Corporation / initiated 2010-05-17 / Terminated / root cause: design

Boston Scientific Corporation began a recall of Ultraflex Esophageal Stent System. Uncovered - Distal Release - 18/23/15. Catalog Number M00513720. Single Use, Rx only. Boston Scientific, Natick, MA 01760... on . The reason FDA records is "Suture related device failures during deployment and repositioning of the stent". As of , the recall status on record is Terminated, and MDM tags the root cause as design.

Recall record

Recall number
Z-1860-2010
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Boston Scientific Corporation
Product
Ultraflex Esophageal Stent System. Uncovered - Distal Release - 18/23/15. Catalog Number M00513720. Single Use, Rx only. Boston Scientific, Natick, MA 01760. Intended for maintaining esophageal luminal patency in esophageal structures caused by intrinsic and/or extrinsic malignant tumors only and occlusion of concurrent esophageal fistulas.
Product code
ESW Prosthesis, Esophageal
Reason for recall
Suture related device failures during deployment and repositioning of the stent
Root cause tag
design
FDA root cause
Device Design
Action
Boston Scientific issued an "Urgent Medical Device Recall - Immediate Action Required" notification dated May 17, 2010 to U.S. Distributors stating that use and distribution of products should cease immediately and return all affected product to the firm for replacement. Distributors were asked to complete and return reply a verification tracking form. For further information, contact Boston Scientific at 1-866-868-4004.
Distribution
Worldwide Distribution - Only US distributed product is affected by recall
Product codes and lots
Lot Numbers: 11737197, 11773676, 11873568, 11930246, 11931993, 12106608, 12123123, 12321371, 12353709, 1242701, 12440418, 12458739, 12601203, 12603432, 12879476, 12898127, 12921104