Titan XL Handpiece Infrared Therapeutic Heating Lamp, Manufactured by Cutera Inc, Brisbane, CA Infrared...

Cutera Inc / initiated 2010-05-08 / Terminated / root cause: design

Cutera Inc began a recall of Titan XL Handpiece Infrared Therapeutic Heating Lamp, Manufactured by Cutera Inc, Brisbane, CA Infrared therapeutic heating lamp, intended to emit energy in... on . The reason FDA records is "Risk of burns-- under certain conditions, devices containing a gold plated end block could produce energy above what is stated on the device setting". As of , the recall status on record is Terminated, and MDM tags the root cause as design.

Recall record

Recall number
Z-0329-2011
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Cutera Inc
Product
Titan XL Handpiece Infrared Therapeutic Heating Lamp, Manufactured by Cutera Inc, Brisbane, CA Infrared therapeutic heating lamp, intended to emit energy in the infrared spectrum to provide topical heating for the purpose of elevating tissue temperature for the temporary relief of minor muscle and joint pain and stiffness.
Product code
ILY Lamp, Infrared, Therapeutic Heating
Reason for recall
Risk of burns-- under certain conditions, devices containing a gold plated end block could produce energy above what is stated on the device setting.
Root cause tag
design
FDA root cause
Device Design
Action
Cutera sent Certified letters sent to all consignees on May 8, 2010. The letter identified the product, the problem, and the action the consignees should take. Consigness were instructed to immediately stop using the handpiece and complete the attached Recall Tracker Form. A second letter was sent to customers who received a replacement handpiecce but did not return the affected devices. All affected products were to be exchanged for new or reworked devices. For questions contact customer service at (888) 428-8372, ext 5561.
Quantity in commerce
1,344 units
Distribution
Worldwide Distribution - USA including AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, NH, NJ, NM, NV, NY, OK, OR, PA, OH, SC, TN, TX, UT, VA, VT, WA, WI, WV, AND WY and the countries of Argentina, Australia, Austria, Barbados, Belgium, Bermuda, Brazil, Bulgaria, Canada, Chile, China, Columbia, Croatia, Cypress, Denmark, Dominican Republic, Egypt, Finland, France, Germany, Greece, Guadeloupe, Hong Kong, India, Indonesia, Ireland, Israel, Italy, Japan, Lle de la Reunion, Mexico, New Zealand, Phillipines, Poland, Puerto Rico, Romania, Russia, Saudi Arabia, Singapore, Slovakia, South Africa, South Korea, Spain, Switzerland, Thailand, Turkey, UAE, and Ukraine.
Product codes and lots
All products manufactured before October 1, 2009 with gold plated end blocks.