Z-1894-2010
Hospira brand Symbiq One-Channel Infuser; List Number: 16026; Distributed by: Hospira, Inc. Morgan Hill, CA...
Abbott Medical / initiated 2010-04-09 / Terminated / root cause: design
Hospira Inc began a recall of Hospira brand Symbiq One-Channel Infuser; List Number: 16026; Distributed by: Hospira, Inc. Morgan Hill, CA and Manufactured by: Hospira Inc, La Aurora... on . The reason FDA records is "Hospira has received numerous customer complaints of the Symbiq infusion pump's failure to detect air-in-line conditions during operation". As of , the recall status on record is Terminated, and MDM tags the root cause as design.
Recall record
- Recall number
- Z-1894-2010
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Abbott Medical
- Product
- Hospira brand Symbiq One-Channel Infuser; List Number: 16026; Distributed by: Hospira, Inc. Morgan Hill, CA and Manufactured by: Hospira Inc, La Aurora, Heredia, Costa Rica. Intended for the delivery of fluids, solutions, drugs, agents, nutritionals, electrolytes, blood and blood products via parenteral, enteral, intravenous, intra-arterial, subcutaneous, epidural or irrigation routes of administration.
- Product code
- FRN Pump, Infusion
- Reason for recall
- Hospira has received numerous customer complaints of the Symbiq infusion pump's failure to detect air-in-line conditions during operation.
- Root cause tag
- design
- FDA root cause
- Device Design
- Action
- Hospira issued a "Urgent: Medical Device Recall Notification" dated June 11, 2010 to update a previously issued Clinical Bulletin. Customers were informed of the issue and instructed on proper mitigation steps to further avoid the issue associated with the affected product. Consignees were asked to complete and return the Reply Form. For further information, contact Hospira Technical Support Operations at 1-800-332-4002, 6AM to 4PM PST, Monday through Friday.
- Quantity in commerce
- 28,485 total
- Distribution
- Worldwide Distribution -- United States (AL, CA, CO, CT, DE, FL, GA, HI, IA, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, NC, ND, NE, NH, NJ, NY, OH, OK, PA, SC, SD, TN, TX, VA and WA), and Australia, Canada, Colombia, Costa Rica, Malaysia, Mexico, Puerto Rico, Singapore and United Arab Emirates.
- Product codes and lots
- All units including List Number Configurations: 16026-04-01/03/04,51/52,53/54, 79/80,81/82,83/84,87/88 and 89/90; 16026-27-89/90; 16026-13-25/26, 53/54.