Z-1897-2010

Lenstar LS 900, Biomicroscope, Slit-Lamp, AC-powered. HAAG-STREIT AG, Switzerland. Used for obtaining ocular...

Haag-Streit USA Inc / initiated 2010-05-19 / Terminated / root cause: software

Haag-Streit USA Inc began a recall of Lenstar LS 900, Biomicroscope, Slit-Lamp, AC-powered. HAAG-STREIT AG, Switzerland. Used for obtaining ocular measurements and performing calculations to assist... on . The reason FDA records is "During the creation of the Lenstar LS 900 software, the ability to perform a manual adjustment of the retinal thickness measurement was not disabled for products targeted for US distribution. The current software...". As of , the recall status on record is Terminated, and MDM tags the root cause as software.

Recall record

Recall number
Z-1897-2010
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Haag-Streit USA Inc
Product
Lenstar LS 900, Biomicroscope, Slit-Lamp, AC-powered. HAAG-STREIT AG, Switzerland. Used for obtaining ocular measurements and performing calculations to assist in determination of the appropriate power and type of Intraocular Lens for implantation after removal of the natural crystalline lens following cataract removal.
Product code
HLF Device, Measuring, Lens Radius, Ophthalmic
Reason for recall
During the creation of the Lenstar LS 900 software, the ability to perform a manual adjustment of the retinal thickness measurement was not disabled for products targeted for US distribution. The current software installed allows the user to manually alter the retinal thickness measurement, which could modify the automatic axial eye length measurement resulting in the selection of an incorrect in
Root cause tag
software
FDA root cause
PMA
Action
HAAG - Streit USA, Inc. issued notification letters to their customers informing them of the problem. They instructed their customer to refrain from adjusting the gates for the measurement of retinal thickness until after the system has been updated. For further information, contact HAAG - Streit USA, Inc. at 1-513-336-7255.
Quantity in commerce
45 units
Distribution
Nationwide Distribution -- Including AL, AZ, CA, FL, ID, KY, LA, MA, MI, MN, MO, NC, NJ, NY, OH, OR, PA, SC, VA, VT, WA and WV.
Product codes and lots
SERIAL Numbers: 223, 224, 225, 284, 285, 316, 317, 319, 320, 322, 323, 324, 325, 327, 330, 331, 332, 333, 364, 365, 366, 368, 369, 377, 378, 379, 381, 384, 385, 386, 405, 407, 408, 409, 411, 412, 413, 415, 416, 452, 454, 460, 461, 490 and 494.