Arrow Percutaneous Sheath Introducer Kit with Integral Hemostasis Valve/Side Port for use with 7.5 Fr...

Arrow International LLC (subsidiary of Teleflex Incorporated) / initiated 2010-06-07 / Terminated / root cause: sterility

Arrow International Inc began a recall of Arrow Percutaneous Sheath Introducer Kit with Integral Hemostasis Valve/Side Port for use with 7.5 Fr. Catheters; Catalog number AJ-09801 on . The reason FDA records is "Sterility not assured. The stopcock handle position was inadvertently changed from an open to a closed position, which prevents sterilization of the closed portion of the device. Sterility of the affected product cannot...". As of , the recall status on record is Terminated, and MDM tags the root cause as sterility.

Recall record

Recall number
Z-2274-2010
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Arrow International LLC (subsidiary of Teleflex Incorporated)
Product
Arrow Percutaneous Sheath Introducer Kit with Integral Hemostasis Valve/Side Port for use with 7.5 Fr. Catheters; Catalog number AJ-09801
Product code
KGZ Accessories, Catheter
Reason for recall
Sterility not assured. The stopcock handle position was inadvertently changed from an open to a closed position, which prevents sterilization of the closed portion of the device. Sterility of the affected product cannot be guaranteed.
Root cause tag
sterility
FDA root cause
Process change control
Action
June 7, 2010 Notification letter to customer instructed them to return the product to Arrow. Arrow will destroy the returned product.
Quantity in commerce
3000 kits
Distribution
All affected product was distributed to one foreign consignee: Arrow Japan Ltd.
Product codes and lots
Lot numbers RF0018145, RF0021641, RF9044164, RF9069739, RF9070837, RF9100262