Arrow Percutaneous Sheath Introducer Kit with Integral Hemostasis Valve/Side Port for use with 7.5 Fr...
Arrow International Inc began a recall of Arrow Percutaneous Sheath Introducer Kit with Integral Hemostasis Valve/Side Port for use with 7.5 Fr. Catheters; Catalog number AJ-09801 on . The reason FDA records is "Sterility not assured. The stopcock handle position was inadvertently changed from an open to a closed position, which prevents sterilization of the closed portion of the device. Sterility of the affected product cannot...". As of , the recall status on record is Terminated, and MDM tags the root cause as sterility.
Recall record
- Recall number
- Z-2274-2010
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Arrow International LLC (subsidiary of Teleflex Incorporated)
- Product
- Arrow Percutaneous Sheath Introducer Kit with Integral Hemostasis Valve/Side Port for use with 7.5 Fr. Catheters; Catalog number AJ-09801
- Product code
- KGZ Accessories, Catheter
- Reason for recall
- Sterility not assured. The stopcock handle position was inadvertently changed from an open to a closed position, which prevents sterilization of the closed portion of the device. Sterility of the affected product cannot be guaranteed.
- Root cause tag
- sterility
- FDA root cause
- Process change control
- Action
- June 7, 2010 Notification letter to customer instructed them to return the product to Arrow. Arrow will destroy the returned product.
- Quantity in commerce
- 3000 kits
- Distribution
- All affected product was distributed to one foreign consignee: Arrow Japan Ltd.
- Product codes and lots
- Lot numbers RF0018145, RF0021641, RF9044164, RF9069739, RF9070837, RF9100262