Beckman Coulter Access Free T3 Calibrators, Part Number A13430 in vitro diagnostic
Beckman Coulter Inc. began a recall of Beckman Coulter Access Free T3 Calibrators, Part Number A13430 in vitro diagnostic on . The reason FDA records is "Beckman Coulter has confirmed reports that the Access Free T3 calibrator lot numbers 989902, 989903, 989904, 989905, 989906 and 989907 will show a decline in performance before reaching their stated expiration date. The...". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-2178-2011
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Siemens Healthcare Diagnostics Inc.
- Product
- Beckman Coulter Access Free T3 Calibrators, Part Number A13430 in vitro diagnostic.
- Product code
- JIT Calibrator, Secondary
- Reason for recall
- Beckman Coulter has confirmed reports that the Access Free T3 calibrator lot numbers 989902, 989903, 989904, 989905, 989906 and 989907 will show a decline in performance before reaching their stated expiration date. The affected calibrator lots will cause one or more quality control samples to produce out-of-range high result.
- Root cause tag
- manufacturing process
- FDA root cause
- Nonconforming Material/Component
- Action
- Beckman Coulter sent a letter dated December 29, 2009, providing the customers with an explanation of the reason for recall and asked them to take the following actions: - Discontinue use of the lots of reagent listed above. - Review your historical quality control to ensure that the assay performance of these lots has been within your laboratory specifications. NOTE: Out-of-range high quality control results may indicate invalid test results. Consignees were asked to share this information with their laboratory staff and retain this notification as part of their laboratory Quality System documentation; and complete and return the enclosed response form within 10 days so that Beckman can be assured that they received this important notification. Assistance or questions regarding this notification, were directed to Technical Support at 1-800-854-3633 in the United States and Canada. Outside the United States and Canada, contact consignees were told to contact their local Beckman Coulter Representative.
- Quantity in commerce
- 2,320 kits
- Distribution
- Distribution Nationwide and Canada.
- Product codes and lots
- Lot #'s: 989902, 989903, 989904, 989905, 989906 & 989907