Hydradjust IV DR Urological Table Sedecal Generator (p/n 750732 - 64W generator & p/n 750733 80W generator)...
Mallinckrodt Inc began a recall of Hydradjust IV DR Urological Table Sedecal Generator (p/n 750732 - 64W generator & p/n 750733 80W generator). Mallinckrodt, Inc. Cincinnati, OH 45237... on . The reason FDA records is "Covidien received complaints of the Sedecal Generator failing and producing smoke at customer sites. The failure modes are located in different parts of the generator and will lead to the generator becoming inoperative". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-1904-2010
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Mallinckrodt, Inc.
- Product
- Hydradjust IV DR Urological Table Sedecal Generator (p/n 750732 - 64W generator & p/n 750733 80W generator). Mallinckrodt, Inc. Cincinnati, OH 45237. Facilitates digital radiologic and/or fluoroscopic procedures requiring a beam of diagnostic quality radiation and a flat imaging table.
- Product code
- IZO Generator, High-Voltage, X-Ray, Diagnostic
- Reason for recall
- Covidien received complaints of the Sedecal Generator failing and producing smoke at customer sites. The failure modes are located in different parts of the generator and will lead to the generator becoming inoperative.
- Root cause tag
- manufacturing process
- FDA root cause
- Process control
- Action
- Covidien Issued an "Urgent Device Correction" notification dated June 14, 2010 informing their customers of the potential problem. Consignees were instructed that Field Service Engineers will be in contact arrange a service visit to perform corrections on the affected units. For further information, contact Covidien Product Monitoring at 1-800-778-7898.
- Quantity in commerce
- 144 units
- Distribution
- Worldwide Distribution -- United States, Canada and China.
- Product codes and lots
- Part Number: 750732, Serial Numbers: G28619, G28622, G28623, G28625, G28627, G28630, G28633, G28668, G28669, G28679, G28680, G28924, G28926, G28928, G28930, G28939, G28997, G29001, G29002, G29006, G29028, G29029, G29063, G29064, G29065, G29068, G29069, G29070, G29071, G29138, G29141, G29143, G29151, G29155, G29157, G29162, G29234, G29236, G29238, G29242, G29243, G29248, G29257, G29258, G29261, G29265, G29268, G29277, G29814, G30632, G30635, G30852, G30856, G31123, G31128, G31134, G31344, G31347, G31355, G31359, G31435, G31437, G31439, G31440, G31503, G31505, G31609, G31610, G31612, G31614, G31616, G32275, G32279, G32284, G32288, G32290, G32456, G32461, G32463, G32464, G32468, G32470, G32476, G32511, G32516, G32518, G32520, G32523, G32524, G32537, G32541, G32565, G32566, G32568, G32571, G32578, G32644, G32647, G32649, G32650, G32774, G32776, G32778, G32779, G32786, G32912, G32914, G32915 and G32922; Part Number: 750733, Serial Numbers: G29230, G29891, G30029, G30171, G30172, G30174, G30180, G30322, G30325, G30661, G30666, G30799, G30801, G31007, G31010, G31307, G31310, G315881, G32292, G32535, G32771, G32773, G32918, G-32924, G-32976, G329821, G33085, G33086, G33090, G33361, G33874, G33876, G34118, G34128 and G34232.