GE Responder 2000 Defibrillator/Monitor, models 2025653, 2026109, and 2026114 that were manufactured between...
Cardiac Science Corporation began a recall of GE Responder 2000 Defibrillator/Monitor, models 2025653, 2026109, and 2026114 that were manufactured between July 25, 2008 and October 2, 2009 by GE Healthcare on . The reason FDA records is "Some units may be non-operational during a rescue, due to a defective PCB interconnect on traces connected to the reed relay. This will result in an inability to analyze the patient's heart rhythm or deliver therapy...". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-0589-2011
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- ZOLL Medical Corporation
- Product
- GE Responder 2000 Defibrillator/Monitor, models 2025653, 2026109, and 2026114 that were manufactured between July 25, 2008 and October 2, 2009 by GE Healthcare.
- Product code
- MKJ Automated External Defibrillators (Non-Wearable)
- Reason for recall
- Some units may be non-operational during a rescue, due to a defective PCB interconnect on traces connected to the reed relay. This will result in an inability to analyze the patient's heart rhythm or deliver therapy during a rescue.
- Root cause tag
- manufacturing process
- FDA root cause
- Nonconforming Material/Component
- Action
- Cardiac Science issued a Urgent Medical Device Corrective Action letter dated April 28, 2010 requesting GE Healthcare cease distribution of affected devices and to instruct end users of distributed devices to schedule a service call to have the device corrected. Cardiac will send rework kits no later than June 30, 2010. CSC can be contacted at 44 161 926 0011 to arrange return material authorization and replacement of affected devices.
- Quantity in commerce
- 130 units
- Distribution
- Worldwide Distribution: Throughout the US, and in the countries of Germany, Korea, Peru, Singapore, and Spain.
- Product codes and lots
- Serial Numbers: 3201281, 3201806, 3201810, 3201815, 3202223, 3202233, 3202244, 3202248, 3202249, 3202253, 3202255, 3202256, 3202257, 3202260, 3202262, 3202269, 3202271, 3202273, 3202275, 3202280, 3202281, 3202283, 3202284, 3202285, 3202286, 3202287, 3202288, 3202290, 3202291, 3202292, 3202297, 3202299, 3202300, 3202301, 3202303, 3202304, 3202305, 3202306, 3202308, 3202310, 3202311, 3202312, 3202313, 3202315, 3202316, 3202317, 3202319, 3202321, 3202324, 3202327, 3202328, 3202331, 3202332, 3202333, 3202334, 3202336, 3202337, 3202338, 3202339, 3202340, 3202344, 3202347, 3202348, 3202349, 3202350, 3202351, 3202352, 3202354, 3202355, 3202356, 3202357, 3202358, 3202360, 3202361, 3202362, 3202363, 3202364, 3202365, 3202371, 3202377, 3202379, 3202380, 3202381, 3202382, 3202385, 3202386, 3202387, 3202388, 3202389, 3202390, 3202391, 3202392, 3202393, 3202394, 3202395, 3202396, 3202397, 3202398, 3202399, 3202400, 3202401, 3202403, 3202404, 3202405, 3202406, 3202407, 3202408, 3202409, 3202411, 3202412, 3202414, 3202415, 3202416, 3202417, 3202418, 3202419, 3202420, 3202421, 3202422, 3202423, 3202427, 3202429, 3202439, 3202442, 3202443, 3202444, 3202447, 3202448, 3202450, and 3202451.