Z-2029-2010

Medtronic StealthStation S7 System

Medtronic Navigation, Inc / initiated 2010-06-14 / Terminated / root cause: design

Medtronic Navigation, Inc began a recall of Medtronic StealthStation S7 System on . The reason FDA records is "Excessive wear at the power cord retractor or incorrect assembly of the power plug may cause sparking, intermittent loss of power, or electrical shock". As of , the recall status on record is Terminated, and MDM tags the root cause as design.

Recall record

Recall number
Z-2029-2010
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Medtronic Navigation, Inc
Product
Medtronic StealthStation S7 System
Product code
HAW Neurological Stereotaxic Instrument
Reason for recall
Excessive wear at the power cord retractor or incorrect assembly of the power plug may cause sparking, intermittent loss of power, or electrical shock.
Root cause tag
design
FDA root cause
Device Design
Action
Customers were notified by letter on Jun 14, 2010, advising them of the problem. Instructions were provided to continue using existing retractable power cords until they could be replaced with non-retractable cords by Medtronic, beginning on Aug 17. Further information is available from the company by calling 800-595-9709.
Quantity in commerce
370 units
Distribution
Nationwide, including 3 VA facilities. No military distribution. Foreign accounts in Australia, Austria, Belgium, Canada, Chile, Czech Republic, Denmark, Egypt, Finland, France, Germany, Greece, Hong Kong, Hungary, India, Indonesia, Ireland, Italy, Latvia, Libya, Lithuania, Luxembourg, Malaysia, Namibia, Netherlands, New Zealand, Poland, Puerto Rico, Russia, Saudi Arabia, Singapore, Slovakia, South Africa, South Korea, Spain, Sweden, Switzerland, Thailand, Turkey, United Kingdom, United Arab Emirates.
Product codes and lots
All lots.