Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-0429-2011

diaDexus PLAC Test ELISA Kit; Manufactured and distributed by diaDexus , Inc., South San Francisco, CA; Model...

Diadexus, Inc. / initiated 2010-03-15 / Terminated

DiaDexus, Inc began a recall of diaDexus PLAC Test ELISA Kit; Manufactured and distributed by diaDexus , Inc., South San Francisco, CA; Model Part Number: 90123 The diaDexusPLAC Test ELISA... on . The reason FDA records is "The reason for the recall is that Real Time Stability Test program showed the kit lot failed stability prior to expiration, affecting samples with results> 500 ng/ml. Results up to 500 ng/ml exhibit passing stability...". As of , the recall status on record is Terminated.

Recall record

Recall number
Z-0429-2011
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Diadexus, Inc.
Product
diaDexus PLAC Test ELISA Kit; Manufactured and distributed by diaDexus , Inc., South San Francisco, CA; Model Part Number: 90123 The diaDexusPLAC Test ELISA Kit is an enzyme Immunoassay for the quantitative determination of Lp-PLA2 (lipoprotein-associated phospholipase A2) in human plasma and serum, to be used in conjunction with clinical evaluation and patient risk assessment as an aid in predicting risk for coronary hear disease, and ischemic stroke associated with atherosclerosis.
Product code
NOE Test, System, Immunoassay, Lipoprotein-Associated Phospholipase A2
Reason for recall
The reason for the recall is that Real Time Stability Test program showed the kit lot failed stability prior to expiration, affecting samples with results> 500 ng/ml. Results up to 500 ng/ml exhibit passing stability results and are not impacted by this issue.
FDA root cause
Other
Action
DiaDexus notified all customers by phone, including distributor's customers, beginning on March 15, 2010. Customers were told to re-test samples that were tested from 2/1/2010 through 2/28/2010 which have values above 500 ng/ml with a new lot of product. For questions regarding this recall call (650) 246-6477.
Quantity in commerce
239 kits
Distribution
Nationwide Distribution - US including NJ, OH, KS, IN, FL, CA, UT, OK, PA, WA and MI.
Product codes and lots
Lot Number: 907090; Part Number: 90123; Product Code NOE