Varian Medical Systems FastPlan software, Versions 5.5 and 5.5.1, Model Number HZ3, manufactured by Varian...
Varian Medical Systems Oncology Systems began a recall of Varian Medical Systems FastPlan software, Versions 5.5 and 5.5.1, Model Number HZ3, manufactured by Varian Medical Systems, Palo Alto, CA Treatment Planning... on . The reason FDA records is "Due to software Issue, an error occurs when performing cone planning using MRI, which may result in the patient receiving a higher radiation dose than expected, potentially to healthy tissue". As of , the recall status on record is Terminated, and MDM tags the root cause as design.
Recall record
- Recall number
- Z-0836-2011
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Varian Medical Systems, Inc.
- Product
- Varian Medical Systems FastPlan software, Versions 5.5 and 5.5.1, Model Number HZ3, manufactured by Varian Medical Systems, Palo Alto, CA Treatment Planning System is intended as aid in quickly and precisely planning the radiation treatment of small cranial lesions such as an asteriovenous malformations, pituitar tumors, pincalomas, acoustic neuromas and malignant neoplasms.
- Product code
- IYE Accelerator, Linear, Medical
- Reason for recall
- Due to software Issue, an error occurs when performing cone planning using MRI, which may result in the patient receiving a higher radiation dose than expected, potentially to healthy tissue.
- Root cause tag
- design
- FDA root cause
- Device Design
- Action
- Varian sent Urgent Medical Device Correction letters on March 23, 2010, by Federal Express to all customers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to run test cases or otherwise determine whether the MR images that they receive have the 'Series Date' parameter populated. Customers should contact their MR system manufacturer to get a confirmation that the 'Series Date' field is populated. Customers were instructed to confirm by visual inspection that the fusion results are loaded appropriately in Cone Planning before starting to create or edit a plan. For questions regarding this recall contact your local Varian Customer Support Distirct or Regional Manager. Phone: USA and Canada - 1-888-VARIANS (888-827-4265) Europe -+41 41 749 8844 E-mail: North America: [email protected] Australia/New Zealand: [email protected] Europe: [email protected] South East Asia: [email protected] China/Asia: [email protected] Japan: [email protected] Latin America: [email protected] Internet: Oncology Systems customer site - www.myvarian.com Varian Medical System public site - www.varian.com
- Quantity in commerce
- 218 units
- Distribution
- Worldwide distribution - USA nationwide and the countries of Australia, Brazil, Canada, Chile, Colombia, Georgia, Germany, Hong Kong, India, Ireland, Japan, Macau, Malaysia, Russia, Singapore, Spain, Sri Lanka, and Taiwan
- Product codes and lots
- All codes start with HZ3: 0001, 0021, 0032, 0033, 0044, 0403, 0404, 0407, 0422, 0426, 0681, 0684, 0698, 0703, 0727, 2035, 2036, 2037, 2039, 2040, 2126, 3020, 3021, 3022, 3026, 9016, 9018, 9066, 9071, 9165, 0054, 0058, 0061, 0077, 0108, 0110, 0111, 0112, 0131, 0139, 0144, 0154, 0167, 0171, 0178, 0182, 0190, 0207, 0208, 0224, 0230, 0251, 0258, 0265, 0273, 0280, 0283, 0293, 0294, 0312, 0330, 0342, 0349, 0368, 0398, 0460, 0461, 0462, 0485, 0495, 0500, 0507, 0508, 0509, 0510, 0511, 0513, 0520, 0531, 0537, 0540, 0556, 0557, 0571, 0572, 0573, 0588, 0594, 0604, 0612, 0617, 0620, 0624, 0630, 0640, 0646, 0649, 0656, 0670, 0675, 0730, 0731, 0737, 0740, 0743, 0762, 2001, 2002, 2003, 2004, 2005, 2006, 2007, 2008, 2009, 2010, 2012, 2013, 2014, 2016, 2017, 2018, 2018, 2020, 2021, 2022, 2024, 2025, 2026, 2027, 2028, 2029, 2032, 2033, 2034, 2042, 2044, 2045, 2046, 2047, 2048, 2049, 2050, 2051, 2052, 2053, 2054, 2055, 2056, 2058, 2059, 2060, 2061, 2062, 2063, 2064, 2065, 2067, 2069, 2071, 2072, 2073, 2074, 2075, 2076, 2077, 2084, 2086, 2108, 2125, 3030, 3031, 3039, 3041, 3042, 3044, 3046, 3047, 3049, 3050, 3053, 3055, 3059, 3062, 3064, 5070, 6101, 6102, 6302, 6303, 7002, 7003, 7005, 7007, 7010, 7011, 7015, 7027, 7101, 8006, 8062, 9002, 9004, 9007, 9010