Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-0013-2011

Summit Broach Handle, NON-STERILE, MFG: DEPUY ORTHOPAEDICS, INC., PO BOX 988 WARSAW, IN 46581, 1 800 366 8143...

Depuy Orthopaedics Inc. / initiated 2010-06-15 / Terminated / root cause: manufacturing process

Depuy Orthopaedics, Inc. began a recall of Summit Broach Handle, NON-STERILE, MFG: DEPUY ORTHOPAEDICS, INC., PO BOX 988 WARSAW, IN 46581, 1 800 366 8143 Catalog Number: 2570-00-000 The Summit Broach... on . The reason FDA records is "Firm discovered that some of their straight broach handles are experiencing early failure due to a weakened handle locking mechanism (Leaf Spring)". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-0013-2011
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Depuy Orthopaedics Inc.
Product
Summit Broach Handle, NON-STERILE, MFG: DEPUY ORTHOPAEDICS, INC., PO BOX 988 WARSAW, IN 46581, 1 800 366 8143 Catalog Number: 2570-00-000 The Summit Broach Handle (instrumentation) is used with the DePuy Summit Tapered Hip System. The Broach Handle is used to prepare the femur for hip system implantation. The broach handle is an impaction/extraction device that holds broaches for creating the cavity in the femur during hip arthroplasty.
Product code
LXH Orthopedic Manual Surgical Instrument
Reason for recall
Firm discovered that some of their straight broach handles are experiencing early failure due to a weakened handle locking mechanism (Leaf Spring).
Root cause tag
manufacturing process
FDA root cause
Process control
Action
On June 3, 2010, an "Alert: Femoral Broach Handle Repair" went out to the entire sales force/customers via "Sales Mail" and posted to the DePuy Internal website. The Alert stated that "We have discovered that some of our straight broach handles are experiencing early failure due to a weakened handle locking mechanism (Leaf Spring). As you perform routine maintenance on your sets, if you have any of these broaches stamped with the lot numbers listed, it is recommended you have them sent in to be reworked with a new spring device." Send the broach handles to: DePuy Orthopaedics Attn: Repairs 700 Orthopaedic Dr Warsaw, IN 46581 Please include return shipping information If you have any questions regarding this information or need additional information, call 574-372-7333.
Quantity in commerce
8358
Distribution
Nationwide distribution: AR, AZ, CA, FL, HI, IA, IL, IN, KS, KY, LA, MA, MD, MI, MN, NC, NV, NY, OH, PA, TN, VA, and WI.
Product codes and lots
Catalog Number: 2570-00-000 Lot Numbers: A1109, A1209, A0210, A0310