Monarch Labs Medical Maggots with the accessory LeFlap dressing (previously known as Creature Comforts II)...
Monarch Labs, LLC began a recall of Monarch Labs Medical Maggots with the accessory LeFlap dressing (previously known as Creature Comforts II), 250-500 per vial, Batch #: MM-090406 with... on . The reason FDA records is "Monarch Labs initiated the field notification regarding their Medical Maggots from Batch MM090406 with the accessory LeFlap dressing due a customer complaint from a physician of a patient developing cellulitis after...". As of , the recall status on record is Terminated.
Recall record
- Recall number
- Z-2207-2010
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Monarch Labs, LLC
- Product
- Monarch Labs Medical Maggots with the accessory LeFlap dressing (previously known as Creature Comforts II), 250-500 per vial, Batch #: MM-090406 with CCII-0901. Maggot therapy is intended to debride serious non-healing (chronic) wounds, and the dressing intended to confine the maggots to the wound.
- Product code
- NQK Maggots, Medical
- Reason for recall
- Monarch Labs initiated the field notification regarding their Medical Maggots from Batch MM090406 with the accessory LeFlap dressing due a customer complaint from a physician of a patient developing cellulitis after using maggots from the affected batch.
- FDA root cause
- Other
- Action
- The customer recall communication was initiated on May 3, 2009 with Monarch Labs notifying all consignees with a via a faxed Recall Letter with fax-back attachment of the initial compliant they had received and the possible infection that may occur if instructions for use are not followed. All therapists, who used maggots from that batch were instructed to contact the firm immediately if they have experienced any adverse events, such as infection, cellulitis, or any other problem. If not, they were asked to simply return the fax-back attachment , noting "no problem" .
- Quantity in commerce
- 14 vials
- Distribution
- Nationwide Distribution -- CA, IA, KY, IL, TX, SC & UT.
- Product codes and lots
- Batch #: MM-090406 with CCII-0901
FIRST ARTICLE / WEEKLY