TomoTherapy Hi-Art System, ver. 4.0.0, 4.0.1, 4.0.2, ThomoTherapy Incorporated 1240 Deming Way, Madison, WI...
TomoTherapy Incorporated began a recall of TomoTherapy Hi-Art System, ver. 4.0.0, 4.0.1, 4.0.2, ThomoTherapy Incorporated 1240 Deming Way, Madison, WI 53717. The TomoTherapy HI-ART System is intended to... on . The reason FDA records is "An issue has been identified, as a result of input received from one TomoTherapy Hi-Art Treatment System customer. With a particular workflow, the Planning Station's Draft and/or final Plan Report can provide incorrect...". As of , the recall status on record is Terminated, and MDM tags the root cause as software.
Recall record
- Recall number
- Z-2072-2010
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Accuray Incorporated
- Product
- TomoTherapy Hi-Art System, ver. 4.0.0, 4.0.1, 4.0.2, ThomoTherapy Incorporated 1240 Deming Way, Madison, WI 53717. The TomoTherapy HI-ART System is intended to be used as an integrated system for the planning and delivery of intensity modulated radiation therapy (IMRT). The HI-ART System provides precise delivery of radiation to tumors or other targeted tissues while minimizing the delivery of radiation to vital healthy tissue.
- Product code
- MUJ System, Planning, Radiation Therapy Treatment
- Reason for recall
- An issue has been identified, as a result of input received from one TomoTherapy Hi-Art Treatment System customer. With a particular workflow, the Planning Station's Draft and/or final Plan Report can provide incorrect dose statistics.
- Root cause tag
- software
- FDA root cause
- Software design
- Action
- Consignees were sent on 6/23/10 a TomoTherapy "Urgent Field Safety Notice Medical device Correction" letter dated June 22, 1010. The letter addressed the Issue, Product Affected, Recommended Action and Resolution. Prior to generating a draft or final Plan Report for a given plan, the plan must be closed and reopened at least once. This issue will be remedied by Hi-Art software version 4.0.3, scheduled for release in late 2010. Customers with questions were provided with a contact in the letter at the TomoTherapy Customer Interaction Center. The firm can be contacted at 608-824-2800.
- Quantity in commerce
- 29
- Distribution
- CA, CO, IL, LA, MI, MO, NY, ND, OH, OK, PA,VA, WI, BELGIUM CANADA, KOREA, MALAYSIA, SWITZERLAND, ITALY, GREAT BRITAIN, FRANCE.
- Product codes and lots
- Serial #'s: 110019, 110083, 110169, 110284, 110305, 110312, 110034, 110091, 110174, 110285, 110306, 110314, 110037, 110092, 110278, 110288, 110308, 110316, 110056, 110150, 110281, 110293, 110309, 110318, 110065, 110159, 110283, 110300, 110311