Boston Scientific, Fathom"-14 Steerable Guidewire, Peripheral Vascular Use, 300 cm x 10 cm, REF 50-814, UPN...
Boston Scientific Corporation began a recall of Boston Scientific, Fathom"-14 Steerable Guidewire, Peripheral Vascular Use, 300 cm x 10 cm, REF 50-814, UPN M001508140 Sterilized with ethylene oxide gas, Rx... on . The reason FDA records is "Boston Scientific is initiating a Medical Device Recall regarding specific lots of the FathomTM-14 Steerable Guidewire. For lubricity, the distal portion of these guidewires is coated with hydrophilic polymer and the...". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-2070-2010
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Boston Scientific Corporation
- Product
- Boston Scientific, Fathom"-14 Steerable Guidewire, Peripheral Vascular Use, 300 cm x 10 cm, REF 50-814, UPN M001508140 Sterilized with ethylene oxide gas, Rx Only, Made in 2405 Orion Circle, West Valley City, UT 84119 USA.
- Product code
- DQX Wire, Guide, Catheter
- Reason for recall
- Boston Scientific is initiating a Medical Device Recall regarding specific lots of the FathomTM-14 Steerable Guidewire. For lubricity, the distal portion of these guidewires is coated with hydrophilic polymer and the proximal portion is coated with polytetrafluoroethylene (PTFE). Through internal inspection process, the firm determined that Fathom guidewires from affected lots may exhibit a P
- Root cause tag
- manufacturing process
- FDA root cause
- Process design
- Action
- Consignees were sent on 6/3/10 an "Urgent Medical Device Recall" letter dated 3 June 2010. The letter was addressed to Risk Manger/Interventional Radiology Manger/Material Manger. The letter described the problem and the product involved. The letter included the "Customer Instructions" which stated to Immediately discontinue use of an segregate recalled product and, requested customers to return the Reply Verification Tracking Form. Questions are directed to the firm at 510-624-2563.
- Quantity in commerce
- 34 units
- Distribution
- AL, AR, CA, CO, CT, FL, GA, ID, IL, KS, KY, LA, MD, MA, MS, NV, NY, NJ, NM, NY, NC, OH, OK, OR, PA, RI, SC, TN, TX, UT, VT,VA, WA, WT.
- Product codes and lots
- B18947