Z-2070-2010

Boston Scientific, Fathom"-14 Steerable Guidewire, Peripheral Vascular Use, 300 cm x 10 cm, REF 50-814, UPN...

Boston Scientific Corporation / initiated 2010-06-03 / Terminated / root cause: manufacturing process

Boston Scientific Corporation began a recall of Boston Scientific, Fathom"-14 Steerable Guidewire, Peripheral Vascular Use, 300 cm x 10 cm, REF 50-814, UPN M001508140 Sterilized with ethylene oxide gas, Rx... on . The reason FDA records is "Boston Scientific is initiating a Medical Device Recall regarding specific lots of the FathomTM-14 Steerable Guidewire. For lubricity, the distal portion of these guidewires is coated with hydrophilic polymer and the...". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-2070-2010
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Boston Scientific Corporation
Product
Boston Scientific, Fathom"-14 Steerable Guidewire, Peripheral Vascular Use, 300 cm x 10 cm, REF 50-814, UPN M001508140 Sterilized with ethylene oxide gas, Rx Only, Made in 2405 Orion Circle, West Valley City, UT 84119 USA.
Product code
DQX Wire, Guide, Catheter
Reason for recall
Boston Scientific is initiating a Medical Device Recall regarding specific lots of the FathomTM-14 Steerable Guidewire. For lubricity, the distal portion of these guidewires is coated with hydrophilic polymer and the proximal portion is coated with polytetrafluoroethylene (PTFE). Through internal inspection process, the firm determined that Fathom guidewires from affected lots may exhibit a P
Root cause tag
manufacturing process
FDA root cause
Process design
Action
Consignees were sent on 6/3/10 an "Urgent Medical Device Recall" letter dated 3 June 2010. The letter was addressed to Risk Manger/Interventional Radiology Manger/Material Manger. The letter described the problem and the product involved. The letter included the "Customer Instructions" which stated to Immediately discontinue use of an segregate recalled product and, requested customers to return the Reply Verification Tracking Form. Questions are directed to the firm at 510-624-2563.
Quantity in commerce
34 units
Distribution
AL, AR, CA, CO, CT, FL, GA, ID, IL, KS, KY, LA, MD, MA, MS, NV, NY, NJ, NM, NY, NC, OH, OK, OR, PA, RI, SC, TN, TX, UT, VT,VA, WA, WT.
Product codes and lots
B18947