Temno Coaxial Chiba Fine Needle Aspiration; a coaxial introducer needle (20G x 15 cm) with a Chiba biopsy...

Medtronic, Inc. / initiated 2010-06-25 / Terminated

Carefusion 2200 Inc began a recall of Temno Coaxial Chiba Fine Needle Aspiration; a coaxial introducer needle (20G x 15 cm) with a Chiba biopsy needle (22G x 20 cm); sterile, for single use only... on . The reason FDA records is "The introducer needle is only 10 cm in length instead of the required 15 cm". As of , the recall status on record is Terminated.

Recall record

Recall number
Z-2111-2010
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Medtronic, Inc.
Product
Temno Coaxial Chiba Fine Needle Aspiration; a coaxial introducer needle (20G x 15 cm) with a Chiba biopsy needle (22G x 20 cm); sterile, for single use only sets, 5 sets per case; CareFusion, McGaw Park, IL 60085 USA, Made in Dominican Republic; REF CCH2220 For use in soft tissue aspiration such as breast, kidney, liver, lung, thyroid, lymph nodes and other various soft tissue masses. It is not intended for use in the bone.
Product code
GDF Guide, Needle, Surgical
Reason for recall
The introducer needle is only 10 cm in length instead of the required 15 cm.
FDA root cause
Mixed-up of materials/components
Action
Cardinal Health, on behalf of CareFusion, telephoned the customers on June 15, 2010 and sent a follow-up "URGENT: PRODUCT RECALL" letter on the same date via UPS next day air. The letter describes product, problem, and action taken by customers. The customers were instructed to immediately stop using affected product, return the enclosed acknowledgment form via fax to 847-686-9101 (note: Cardinal Health is required to confirm receipt of this notification from the customers), notify any customers to whom they may have distributed product to, contact CareFusion Customer Service at 800-653-6827 to arrange for credit and return of the affected product, and send the affected product to CareFusion, 1240 Waukegan Rd, Waukegan, IL 60085, Attn: Scott Fitzer, after obtaining an RMA/RGA number from CareFusion. If you have any questions, please contact 800-292-9332.
Quantity in commerce
65 units
Distribution
Nationwide distribution: California, Florida, Georgia, Illinois, Kentucky, Massachusetts, New Jersey, New York and Pennsylvania.
Product codes and lots
Catalog number CCH2220, Lot D09121745