Z-2112-2010

IMPAX CV Results Manager/Results Manager Administration Tool

AGFA Healthcare Corp. / initiated 2010-06-11 / Terminated

AGFA Corp. began a recall of IMPAX CV Results Manager/Results Manager Administration Tool on . The reason FDA records is "Agfa Service technician had incorrectly modified a report template at one site resulting in erroneous anatomic segment locators". As of , the recall status on record is Terminated.

Recall record

Recall number
Z-2112-2010
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
AGFA Healthcare Corp.
Product
IMPAX CV Results Manager/Results Manager Administration Tool
Product code
LLZ System, Image Processing, Radiological
Reason for recall
Agfa Service technician had incorrectly modified a report template at one site resulting in erroneous anatomic segment locators.
FDA root cause
Other
Action
A letter was sent to the consignee on 6/11/2010 indicating what steps would be taken by Agfa concerning the problem. The letter also included a feedback form which was to be returned to Agfa. The recall was expanded and a "URGENT SAFETY NOTICE" was sent to each additional consignee on January 25, 2012. The letter described the potential issue and mitigation. Acknowledgment, via FAX-Back or email, that the information was received and understood has been requested from the additional consignees.
Quantity in commerce
358
Distribution
AK, AZ, CA, CO, DC, FL, GA, HI, IL, IN, IA, KS, KY, LA, ME, MD, MA, MI, MN, MO, MS, MT, NH, NJ, NY, NC, OH, OK, OR, PA, RI, SC, SD, TN, TX, VT, VA, WA, WV, WI and Canada
Product codes and lots
Model No. L9M2100; Software Versions RM 2.08.12.00.00. All versions of the CV Reporting (RM) are now affected.