BIOMET MICROFIXATION REF: 02-0461 REES INSULATED FORCEPS, 21CM, (8.25") (L), STAINLESS STEEL, NYLON...
Biomet Microfixation, Inc. began a recall of BIOMET MICROFIXATION REF: 02-0461 REES INSULATED FORCEPS, 21CM, (8.25") (L), STAINLESS STEEL, NYLON, ***NON-STERILE PRODUCT***. The Rees Insulated Forceps are... on . The reason FDA records is "The forceps insulation may have cracks or breaks in the insulation that may cause burns to patients when used". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-2392-2010
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Biomet Microfixation
- Product
- BIOMET MICROFIXATION REF: 02-0461 REES INSULATED FORCEPS, 21CM, (8.25") (L), STAINLESS STEEL, NYLON, ***NON-STERILE PRODUCT***. The Rees Insulated Forceps are used during surgery to grasp, dissect and coagulate selected tissue. Distributed by: BIOMET MICROFIXATION, 1520 TRADEPORT DRIVE, JACKSONVILLE, FL 32218, www.biometmicrofixation.com
- Product code
- GEN Forceps, General & Plastic Surgery
- Reason for recall
- The forceps insulation may have cracks or breaks in the insulation that may cause burns to patients when used.
- Root cause tag
- manufacturing process
- FDA root cause
- Manufacturing material removal
- Action
- Biomet Microfixation issued a letter dated July 2010 to their consignees. Each consignee was ask to discontinue use of the affected product and return the product back to the company for credit. Biomet asked their customers to complete a "Inventory Reconciliation" form and fax the completed form to Biomet Microfixation. If the consignee has further distributed the product, the consignee is asked to contact their customers. Biomet Microfixation can be contacted at 904 741-4400, ext 468.
- Quantity in commerce
- 130 units
- Distribution
- Worldwide Distribution: USA including states of CA, FL, GA, ID, IL, IN, MA, MD, MS, NC, NH, NJ, NV, NY, OH, PA, TX, VA, WA, and WI, and countries of Norway, Canada, and Mexico.
- Product codes and lots
- Lot Number: 042308D08, 112108K08, 031309C09, 041009D09, 042310B10, 021910B10, 092707I07, 102607J07, 032607C07, and 041607D07.