Z-2073-2010

Mast Roller Pump, Catalog 50-80-00, for use with Sorin HLM C5 and HLM S5

LivaNova USA, Inc. / initiated 2010-06-18 / Terminated

Sorin Group USA, Inc. began a recall of Mast Roller Pump, Catalog 50-80-00, for use with Sorin HLM C5 and HLM S5 on . The reason FDA records is "Although accessories were designed for use with two models, the HLM C5 and the HLM S5, the submission that included references to the Mast Roller Pump, B-Care5, and the Electrical Venous Line Occluder accessories...". As of , the recall status on record is Terminated.

Recall record

Recall number
Z-2073-2010
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
LivaNova USA, Inc.
Product
Mast Roller Pump, Catalog 50-80-00, for use with Sorin HLM C5 and HLM S5.
Product code
DTQ Console, Heart-Lung Machine, Cardiopulmonary Bypass
Reason for recall
Although accessories were designed for use with two models, the HLM C5 and the HLM S5, the submission that included references to the Mast Roller Pump, B-Care5, and the Electrical Venous Line Occluder accessories covered the new HLM S5 system. The three accessories were not cleared for use with the HLM C5.
FDA root cause
PMA
Action
The two customers who had received the suspect device accessories were identified and were contacted by letter on 6/18/2010. They were told to immediately discontinue use of the accessories and arrange for return of the products. They were also told that they could continue to use the CD5 Machine without the recalled accessories. A Product Recall Response Form was to completed by each consignee. Questions are directed to the firm at 800-221-7943.
Quantity in commerce
1 unit
Distribution
CA, PA, TN.
Product codes and lots
Serial number: 50K00693