Z-2105-2010
Merit Laureate Hydrophylic Guide Wire, Angled Stiff Shaft, Made In Ireland, CAT No: LWSTFA35260EX, 260cm...
Merit Medical Systems, Inc. / initiated 2010-06-28 / Terminated / root cause: manufacturing process
Merit Medical Systems, Inc. began a recall of Merit Laureate Hydrophylic Guide Wire, Angled Stiff Shaft, Made In Ireland, CAT No: LWSTFA35260EX, 260cm, Sterile EO, Merit Medical Systems, Inc., South... on . The reason FDA records is "Guidewires have decreased lubricity at the proximal end, inhibiting proper operation". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-2105-2010
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Merit Medical Systems, Inc.
- Product
- Merit Laureate Hydrophylic Guide Wire, Angled Stiff Shaft, Made In Ireland, CAT No: LWSTFA35260EX, 260cm, Sterile EO, Merit Medical Systems, Inc., South Jordan, UT 84095. Intended to facilitate the placement of devices during diagnostic and interventional procedures.
- Product code
- DQX Wire, Guide, Catheter
- Reason for recall
- Guidewires have decreased lubricity at the proximal end, inhibiting proper operation.
- Root cause tag
- manufacturing process
- FDA root cause
- Process control
- Action
- The firm, Merit Medical System, Inc., notified the Sales representatives by e-mail on June 28, 2010, and instructed them to contact all of their accounts and to advise them to discontinue use and return any unused products to Merit. A Product Recall Notice was to be completed by each facility under the direction of the rep and all activities were to be completed by July 9, 2010. Questions should be directed to Penny Goldman, Senior Product Manager at 801-208-4198 or Jackie Preece, Manager, Customer Service at 801-208-4365.
- Quantity in commerce
- 140 units
- Distribution
- Nationwide distribution: CA, CO, FL, GA, IL, IN, KS, MA, MD, MI, MN, NC, NH, NJ, NY, PA, and SD.
- Product codes and lots
- Lot Number K131835