Z-2219-2010

Hi-Art System, H-0000-0003 Usage: The TomoTherapy Hi-Art System is intended to be used as an integrated...

Accuray Incorporated / initiated 2009-12-01 / Terminated / root cause: software

TomoTherapy Incorporated began a recall of Hi-Art System, H-0000-0003 Usage: The TomoTherapy Hi-Art System is intended to be used as an integrated system for the planning and precise delivery of... on . The reason FDA records is "An issue was identified with the TomoTherapy HI-Art System. In the event a patient or DQA plan has a moved image, roll adjustments applied during registration will be incorrect. The Planning Station Plan Settings and...". As of , the recall status on record is Terminated, and MDM tags the root cause as software.

Recall record

Recall number
Z-2219-2010
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Accuray Incorporated
Product
Hi-Art System, H-0000-0003 Usage: The TomoTherapy Hi-Art System is intended to be used as an integrated system for the planning and precise delivery of radiation therapy, stereotactic radiotherapy, or stereotactic radiosurgery to tumors or other targeted tissues while minimizing the delivery of radiation to vital healthy tissue. The megavoltage x-ray radiation is delivered in a rotational, non-rotational, modulated (IMRT), or non modulated (non-IMRT/three dimensional conformal) format in accordance with the physician approved plan
Product code
MUJ System, Planning, Radiation Therapy Treatment
Reason for recall
An issue was identified with the TomoTherapy HI-Art System. In the event a patient or DQA plan has a moved image, roll adjustments applied during registration will be incorrect. The Planning Station Plan Settings and DQA Setup tabs allow for images to be moved during planning. During registration when roll is applied on moved images, the Operator Station incorrectly rolls the image about
Root cause tag
software
FDA root cause
Software design
Action
Consignee was sent a TomoTherapy "Urgent Medical device Correction Field Safety Notice' dated December 01, 2009. The letter described the Issue, Product Affected, Recommended Action and Resolution.
Quantity in commerce
1
Distribution
VA
Product codes and lots
110019