Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-2107-2010

GEM Check Coag CNTRL Normal APTT Whole Blood Controls Package Insert: IVD. 15/pk. 06260061600. Manufactured...

Accriva Diagnostics, Inc. / initiated 2010-07-08 / Terminated / root cause: labeling and UDI

International Technidyne Corporation began a recall of GEM Check Coag CNTRL Normal APTT Whole Blood Controls Package Insert: IVD. 15/pk. 06260061600. Manufactured for: Instrumentation Laboratory Company, Lexington... on . The reason FDA records is "Certain lots of GEM Check Coag Whole Blood Controls contain an error in the package insert where the incorrect Acceptable Performance Ranges were published". As of , the recall status on record is Terminated, and MDM tags the root cause as labeling and UDI.

Recall record

Recall number
Z-2107-2010
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Accriva Diagnostics, Inc.
Product
GEM Check Coag CNTRL Normal APTT Whole Blood Controls Package Insert: IVD. 15/pk. 06260061600. Manufactured for: Instrumentation Laboratory Company, Lexington, MA 02421-3125. Intended to be used to perform quality control assays using GEM test cartridges on the IL GEM PCL/PCL Plus Portable Coagulation Laboratory.
Product code
GGN Plasma, Coagulation Control
Reason for recall
Certain lots of GEM Check Coag Whole Blood Controls contain an error in the package insert where the incorrect Acceptable Performance Ranges were published.
Root cause tag
labeling and UDI
FDA root cause
Error in labeling
Action
Urgent: Medical Device Correction Notices, dated July 1, 2010, were sent by Federal Express on July 8, 2010 to all first consignee end users. The letters identified the affected product and asked customers to review and follow all instructions as outlined in the ITC in the communication. The info should be shared with laboratory staff and the notification should be retained as part of the laboratory Quality System documentation. The enclosed response form should be completed and returned to ITC within 5 days.
Quantity in commerce
95 kits C9GC5001; 32 kits G9GC5003; 46 kits J9GC5004; 5 kits C0GC5001; Each kit contains 15 vials.
Distribution
Worldwide Distribution -- USA, including states of NY, TN, CA, AR, IL and TX and countries of Australia, Italy, Spain, Germany, Chile, Spain, Malta, Belgium, Austria, and Switzerland.
Product codes and lots
Part Number 06260061600; Lot C9GC5001, Exp. 7/31/2010; G9GC5003, Exp 1/31/2011; J9GC5004, Exp 3/31/2011; and C0GC5001, Exp 10/31/2011.