Spider Flex Monopolar Hook (Spider Surgical Instruments), TransEnterix, Inc., Durham, NC

Asensus Surgical, Inc. / initiated 2010-06-22 / Terminated

Transenterix, INC began a recall of Spider Flex Monopolar Hook (Spider Surgical Instruments), TransEnterix, Inc., Durham, NC on . The reason FDA records is "The hook end effecter from a monopolar device may become dislodged and fall into the patient when in use during laproscopic surgery. Retrieval from the patient may involve extended anesthesia time". As of , the recall status on record is Terminated.

Recall record

Recall number
Z-2465-2010
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Asensus Surgical, Inc.
Product
Spider Flex Monopolar Hook (Spider Surgical Instruments), TransEnterix, Inc., Durham, NC
Product code
GCJ Laparoscope, General & Plastic Surgery
Reason for recall
The hook end effecter from a monopolar device may become dislodged and fall into the patient when in use during laproscopic surgery. Retrieval from the patient may involve extended anesthesia time.
FDA root cause
Other
Action
TransEnterix issued a "Field Action" letter dated June 22, 2010 to consignees, identifying the affected product and actions to be taken by the customer. Customers were instructed to return the product to TransEnterix, Inc. using the address label and return information provided. The firm will replace the product at no charge. TransEnterix Customer Care can be contacted at 1-888-879-4111.
Quantity in commerce
46 boxes (6 units per box)
Distribution
Nationwide Distribution in the states of AZ, FL, GA, NC, NY, OH, OR, SC and TX.
Product codes and lots
Model number: 9000031, The following lot numbers recalled on June 22, 2010: DSC077201001 and DSC083201001. Field action extended and the following lot numbers recalled on July 5, 2010: DSC043201001, DSC049201001, DSC079201001, DSC091201002