Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-2446-2010

Disposable Detachable Monopolar Needle Electrode for electromyograph model number 110-725, for Physician use...

Chalgren Enterprises, Inc. / initiated 2007-04-12 / Terminated / root cause: packaging

Chalgren Enterprises Inc began a recall of Disposable Detachable Monopolar Needle Electrode for electromyograph model number 110-725, for Physician use, manufactured by Chalgren Enterprises Inc... on . The reason FDA records is "Sterility Compromised-- Investigation found that individual packaging seals were separating, which may compromise sterility". As of , the recall status on record is Terminated, and MDM tags the root cause as packaging.

Recall record

Recall number
Z-2446-2010
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Chalgren Enterprises, Inc.
Product
Disposable Detachable Monopolar Needle Electrode for electromyograph model number 110-725, for Physician use, manufactured by Chalgren Enterprises Inc., Gilroy, CA.
Product code
IKT Electrode, Needle, Diagnostic Electromyograph
Reason for recall
Sterility Compromised-- Investigation found that individual packaging seals were separating, which may compromise sterility.
Root cause tag
packaging
FDA root cause
Packaging
Action
Chalgren Enterprises, Inc. removed all of the affected inventory from inventory and had all affected inventory retrieved from distribution.
Distribution
Nationwide Distribution: USA states of CA, IL, NJ, OH and TX.
Product codes and lots
Lot number P059