CERCLAGE WIRE, 0.6 MM DIA. X 175 MM LENGTH, SST, REF 71163006, QTY: (6), STERILE R, CE 008, 2020/07, Smith &...
Smith & Nephew Inc began a recall of CERCLAGE WIRE, 0.6 MM DIA. X 175 MM LENGTH, SST, REF 71163006, QTY: (6), STERILE R, CE 008, 2020/07, Smith & Nephew, Inc., Memphis, TN 38116, EC REP: Smith &... on . The reason FDA records is "Plastic trays which encase these products have burrs on the hinged packaging which can puncture the single barrier sterile pouch, compromising sterility". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-2197-2010
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Smith & Nephew, Inc.
- Product
- CERCLAGE WIRE, 0.6 MM DIA. X 175 MM LENGTH, SST, REF 71163006, QTY: (6), STERILE R, CE 008, 2020/07, Smith & Nephew, Inc., Memphis, TN 38116, EC REP: Smith & Nephew Orthopaedics GmbH, Tuttlingen, Germany Usage: Bone fracture fixation
- Product code
- JDQ Cerclage, Fixation
- Reason for recall
- Plastic trays which encase these products have burrs on the hinged packaging which can puncture the single barrier sterile pouch, compromising sterility.
- Root cause tag
- manufacturing process
- FDA root cause
- Process control
- Action
- All affected Smith & Nephew Sales Representatives were notified of problem and the recall was sent via overnight mail on 01/23/2009. All affected Smith & Nephew International Distributors were notified via email and telephone on 01/23/2009. They were instructed to immediately quarantine the product for return.
- Quantity in commerce
- 1 unit
- Distribution
- KY, TX, DC, NC, CA, AZ, GA and OH; 3 International: Australia, Malaysia, and Switzerland
- Product codes and lots
- Lot Number: 10DM13784